JACC: Cardiovascular Interventions
Gluing of an Aortic Perforation During Transcatheter Aortic Valve ReplacementAn Alternative Treatment for Annular Rupture?
Author + information
- Received August 5, 2015
- Accepted August 13, 2015
- Published online December 28, 2015.
Author Information
- Nicolas Piliero, MSc∗∗ (npiliero{at}chu-grenoble.fr),
- Fréderic Thony, MD∗,†,
- Gerald Vanzetto, MD, PhD∗,‡ and
- Gilles Barone-Rochette, MD, PhD∗,‡∗ (gbarone{at}chu-grenoble.fr)
- ∗Department of Cardiology, University Hospital, Grenoble, France
- †Department of Radiology, University Hospital, Grenoble, France
- ‡INSERM, U1039, Radiopharmaceutiques Biocliniques, Grenoble, France
- ↵∗Reprint requests and correspondence:
Dr. Nicolas Piliero OR Dr. Gilles Barone-Rochette, Department of Cardiology, University Hospital–BP 217 X, 38043 Grenoble Cedex 9, France.
A 96-year-old man was referred to our cardiology center for a transcatheter aortic valve replacement (TAVR). Pre-implantation transthoracic echocardiography showed preserved ejection fraction, mean transvalvular pressure gradient at 46 mm Hg, aortic valve area of 1 cm2, and maximal aortic jet velocity at 4.1 m/s, confirming severe aortic stenosis. Aortic annulus diameter was 25 mm as measured by transesophageal echocardiography, and its surface was 671 mm2 by 3-dimensional multidetector computed tomography (Figures 1A and 1B). A 29-mm Sapiens XT valve (Edwards Lifesciences, Irvine, California) was implanted using left transfemoral access. During prosthetic valve deployment, contrast extravasation into the pericardium occurred (Figure 1C) causing cardiac shock and pericardial tamponade demonstrated by transthoracic echocardiography. Percutaneous pericardiocentesis and fluid infusion were immediately performed allowing hemodynamic stabilization. Tissue glue composed of a 1/1 mixture of N-butyl-2-cyanoacrylate (Histoacryl) and Lipiodol (B. Braun, Melsungen, Germany) was injected into the perforation using a Progreat microcatheter (Terumo Corporation, Tokyo, Japan) introduced through a 6-F Amplatz left guiding catheter (Medtronic, Dublin, Ireland) (Figure 1D). This successfully stopped the pericardial effusion (Figure 1E) and allowed the procedure to be completed without further complications. The patient was discharged 5 days after the valve deployment.
Computed Tomography and Angiography Imagery Before and During TAVR Procedure
(A and B) Pre-operative computed tomography showed massive calcification on the aortic valve annulus and left anterior leaflet (yellow arrow). (C) A severe contrast extravasation (yellow arrows) was identified during angiography after transcatheter aortic valve implantation. (D) Tissue glue was injected into the perforation with a microcatheter. (E) Severe contrast extravasation was stopped successfully by liquid embolization mixtures.
Aortic rupture occurs in about 1% of all TAVR implantations (1,2). Potential causes include radial forces during valve deployment inducing tissue overdistention along with calcification of the annulus and/or valve leaflets (1). Depending on the severity and location of the perforation, treatment can include: cardiac surgery, isolated percutaneous pericardiocentesis, or a conservative strategy (1). N-butyl-2-cyanoacrylate is a tissue glue that has the capacity to harden immediately when in contact with blood. The adjunction of oily radiopaque substance (Lipiodol) delays polymerization and allows radiopacity (3). Thus, reported here is a novel potential treatment for aortic rupture during TAVR by an endovascular approach.
Acknowledgment
The authors thank Raphaëlle-Ashley Guerbaai for editorial support.
Footnotes
All authors have reported that they have no relationships relevant to the contents of this paper to disclose.
- Received August 5, 2015.
- Accepted August 13, 2015.
- 2015 American College of Cardiology Foundation
References
- ↵
- Pasi M.,
- Unbehaun A.,
- Buz S.,
- Drews T.,
- Hetzer R.
- ↵
- Möllmann H.,
- Kim W.K.,
- Kempfert J.,
- Walther T.,
- Hamm C.
- ↵