Post-Implantation Repositioning of the CoreValve Percutaneous Aortic Valve
Azeem Latib, MD*,
Iassen Michev, MD*,
Jean-Claude Laborde, MD ,
Matteo Montorfano, MD*,
Antonio Colombo, MD*, ,*
* Interventional Cardiology Unit, San Raffaele Scientific Institute, Milan, Italy
Cardiology Department, Glenfield Hospital, Leicester, United Kingdom
Interventional Cardiology Unit, EMO-GVM Centro Cuore Columbus, Milan, Italy
A 76-year-old man presented with congestive heart failure on the basis of severe aortic stenosis (AS). Echocardiography confirmed severe calcific AS (mean gradient = 45 mm Hg, peak gradient = 78 mm Hg, aortic valve area = 0.5 cm2, annulus diameter = 25 mm) and globally depressed left ventricular (LV) function (ejection fraction = 30%). He was considered high-risk for surgical valve replacement (Logistic Euroscore = 37.8%) and referred for transfemoral percutaneous aortic valve implantation. A 29-mm self-expanding CoreValve ReValving System (CoreValve Inc., Irvine, California) prosthesis was implanted retrogradely (1). Low suboptimal prosthesis placement was confirmed angiographically (Fig. 1A) and by the presence of hemodynamically severe acute aortic regurgitation (AR) (Fig. 1B) (2). The upper part of the prosthesis was snared (Fig. 2A) with a 30-mm Amplatz GooseNeck (ev3 Inc., Plymouth, Minnesota). Gentle, continuous traction was applied to the prosthesis with the gooseneck for up to 5 min until proximal displacement visually occurred during fluoroscopy (Fig. 2B). Pulling on the prosthesis with jerky and jarring movements should be avoided. Aortography (Fig. 3A) and hemodynamic monitoring (Fig. 3B) confirmed reduction in AR severity. At 6-month follow-up, the patient was asymptomatic, and echocardiography showed recovery of LV function (ejection fraction = 50%) and residual moderate AR.

View larger version (31K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 1 Aortography and Pressure Tracings Immediately After CoreValve Implantation
(A) Post-implantation aortography documenting low implantation of the CoreValve relative to the aortic annulus (broken line), resulting in severe acute aortic regurgitation. (B) Simultaneous aortic (blue) and left ventricular (red) pressure tracings showing a wide aortic pulse pressure (±85 mm Hg) and equalization of diastolic pressures in late diastole confirming severe aortic regurgitation.
|
|

View larger version (30K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 3 Final Aortography and Pressure Tracings After Repositioning the CoreValve
Final aortography (A) demonstrating a better CoreValve position in relation to the native aortic annulus (broken line) with reduction of the severity of aortic regurgitation, which was confirmed hemodynamically (B) by the finding of a pressure difference in late diastole between the aorta (blue) and left ventricle (red).
|
|
Low implantation of the CoreValve might result in severe acute AR. This case highlights the importance of hemodynamic assessment of post-procedural AR severity in addition to angiographic evaluation of the paravalvular leak in evaluating the LV tolerance of the regurgitation. The experience with transcatheter aortic valve implantation does not differ from balloon aortic valvuloplasty where grade III AR could be well-tolerated in the presence of LV hypertrophy, whereas grade II AR might be poorly tolerated if LV function is depressed. Repositioning of the CoreValve by snaring is feasible and might reduce the severity of AR. However, this high-risk maneuver should be performed cautiously, because it might result in prosthesis embolization into the ascending aorta. If the operator is concerned about the risk of valve embolization, implanting an Edwards-Sapien valve (Edwards Life Sciences, Irvine, California) or another CoreValve inside the first with the valve-in-valve technique could be considered an alternative procedure.
 |
Footnotes
|
|---|
Dr. Laborde is a consultant for and stockholder of CoreValve, Inc.
* Reprint requests and correspondence: Dr. Antonio Colombo, EMO-GVM Centro Cuore Columbus, 48 Via M. Buonarroti, 20145 Milan, Italy (Email: info{at}emocolumbus.it).
Manuscript received June 11, 2009;
revised manuscript received July 21, 2009,
accepted July 25, 2009.
 |
REFERENCES
|
|---|
- Grube E, Schuler G, Buellesfeld L, et al. Percutaneous aortic valve replacement for severe aortic stenosis in high-risk patients using the second- and current third-generation self-expanding CoreValve prosthesis: device success and 30-day clinical outcome J Am Coll Cardiol 2007;50:69-76.[Abstract/Free Full Text]
- Feldman T, Grossman W. Profiles in valvular heart diseaseIn: Baim DS, editor. Grossman's Cardiac Catheterization, Angiography, and Intervention. 7th edition. Philadelphia, PA: Lippincott Williams and Wilkins; 2006. pp. 638-658.
Related Article
-
CoreValve Aortic Bioprosthesis: Repositioning Techniques
- Manolis Vavuranakis, Dimitrios Vrachatis, and Christodoulos Stefanadis
J. Am. Coll. Cardiol. Intv. 2010 3: 565.
[Full Text]
[PDF]
This article has been cited by other articles:

|
 |

|
 |
 
M. Vavuranakis, D. Vrachatis, and C. Stefanadis
CoreValve Aortic Bioprosthesis: Repositioning Techniques
J. Am. Coll. Cardiol. Intv.,
May 1, 2010;
3(5):
565 - 565.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. Latib and A. Colombo
Reply
J. Am. Coll. Cardiol. Intv.,
May 1, 2010;
3(5):
565 - 566.
[Full Text]
[PDF]
|
 |
|
|