Temporal Course of Neointimal Formation After Drug-Eluting Stent PlacementIs Our Understanding of Restenosis Changing?*
Aloke V. Finn, MD ,
Gaku Nakazawa, MD ,
Frank D. Kolodgie, PhD ,
Renu Virmani, MD ,*
CVPath Institute, Inc., Gaithersburg, Maryland
Emory University, Atlanta, Georgia
Key Words: drug-eluting stents polymer-free, restenosis late catch-up
Drug eluting stents (DES) were designed with the primary purpose of inhibiting neointimal formation after percutaneous coronary intervention. Early data from randomized clinical studies confirmed the efficacy of these devices in preventing restenosis when compared with bare-metal stents (BMS) 6 to 8 months after implantation (1,2). However, given that the period of drug release is relatively short (i.e., <90 days) after DES implantation in animal models and that preclinical studies done by us and others demonstrate a late "catch-up phenomenon" in terms of an increase in neointimal thickness from 1 to 3 months, an important unanswered question remains: "Do DES halt the restenotic process or merely delay it?" (3–5). Our data from human pathology from patients dying after DES placement indicate incomplete healing as long as 50 months after device placement (6).
In this issue of JACC: Cardiovascular Interventions, Byrne et al. (7) present data from an observational study of patients receiving permanent polymer rapamycin-eluting stents (RES), permanent polymer paclitaxel-eluting stents (PES), or polymer-free rapamycin-eluting stents (pf-RES) who underwent serial angiographic follow-up at 6 to 8 months (i.e., early) and 2 years (i.e., late) after stent placement. The RES and PES demonstrated significant increases in late lumen loss from early to late follow-up, whereas pf-RES showed essentially no change. These data demonstrate that the arterial response to these devices is a dynamic process lasting well beyond the 6-month time point chosen for angiographic follow-up in clinical trials and highly dependent on the type of DES (i.e., polymeric vs. nonpolymeric) as well as the duration of drug release. These factors might heavily influence the temporal course of intimal formation and healing after DES placement. This study also puts in doubt the belief that polymeric slow release of the drug is essential for long-term efficacy of DES.
Much of our current understanding of the vascular responses to DES has been generated from animal data. Given the limited resolution of conventional angiography, intravascular ultrasound, or other modalities, preclinical histological studies remain the most effective means of evaluating arterial responses to vascular stent implants and have helped lend insights into human responses (8). Although arterial repair after stent placement in normal animal arteries occurs more rapidly than in humans, the sequence of biological responses are remarkably similar (9). Both RES and PES differ from pf-RES in that they contain durable polymers, which serve to extend the duration of drug release but remain as permanent implants capable of provoking their own inflammatory responses in the arterial wall. In contrast, pf-RES contains no polymer but instead consists of 316L stainless steel microporous surface onto which rapamycin is directly sprayed (10).
In animal models, the duration of drug release of the 3 systems is very different. Wessely et al. (10) demonstrated in the porcine model that pf-RES elute two-thirds of their drug in the first week and nearly all of the loaded dose by 21 days, whereas RES stents elute only 68.4% at 28 days (3,10). Moreover, tissue levels peak at 3 days with pf-RES as compared with 14 days with the RES. The temporal release of PES is quite different with a slow release occurring over at least 28 days with the rest "sequestered within the polymer."
These release kinetics translate into very different arterial histologic responses. Although the initial response to polymeric RES and PES is a significant decrease in neointimal growth at 28 days compared with BMS, both stents demonstrate increases in intimal formation beyond this time point such that the initial antirestenotic benefit is lost, which was also reinforced by clinical studies (11,12) (Fig. 1). Although no long-term data exist in animal models for pf-RES, responses resemble that of the BMS with similar amounts of fibrin deposition, inflammation, and endothelialization at 28 days in the porcine and rabbit models (10). These data raise the question of whether the biological effect of pf-RES exists or not. In general, both in humans and in animals, BMS results in an early peak in intimal formation with regression of neointima thereafter (13,14). There is no reason to suspect that the response to pf-RES would be any different than BMS, as was seen histologically in our laboratory.

View larger version (25K):
[in this window]
[in a new window]
[Download PPT slide]
|
Figure 1 Long-Term Outcomes in Porcine Coronary Artery Model and Human Clinical Trials
(A) The bar chart shows neointimal thickness in porcine coronary artery after Cypher and Taxus (overlapping) implantation. Note "late catch-up" from 28 to 90 days in drug-eluting stent, whereas in bare metal stent (BMS) no such increase is observed (14). (B) Bar graph showing clinical follow-up of patients implanted with Cypher and Taxus stent from 1 to 5 years in the SIRIUS and TAXUS IV trials, respectively (11,12). Similarly, target lesion revascularization (TLR) is increasing progressively with each year.
|
|
Two factors are likely responsible for the differing pattern of neointimal growth seen in polymeric DES versus pf-RES. These are differences in duration of drug release and inflammatory responses to polymer. The short duration of drug release in pf-RES essentially translates into BMS-like responses with complete healing at 28 days in animal models. In the rabbit model polymeric RES demonstrate significantly increased fibrin and decreased endothelialization when compared with BMS. These data indicate that, for polymeric DES, arterial healing is incomplete at 28 days and that unlike BMS arterial repair continues beyond this time point. Therefore it is not surprising that, in humans, changes in neointimal growth occur far beyond the normal time point at which healing is complete in BMS (i.e., 6 months).
In addition, the durable polymers on RES (polyethylene-co-vinyl acetate [PEVA] and poly n-butyl methacrylate [PBMA]) and PES (poly[styrene-b-isobutylene-b-styrene]) provide an additional factor that influences local responses and might alter processes involved in neointimal formation. Each polymer provokes a distinctive inflammatory response in animals. In an overlapping rabbit model of stent deployment, we previously reported that RES implants were characterized by giant cell infiltration around stent struts, whereas PES provoked a heterophil/eosinophilic reaction (15). The porcine model demonstrates a progressive granulomatous and eosinophilic reaction to Cypher stent starting at 28 days and increasing out to 1 year (14). The likely explanation for these findings is a local hypersensitivity reaction to the nonerodable polymers used on the Cypher stents (PEVA and PBMA), because it peaks only after the complete release of drug (i.e., >60 days). Carter et al. (4) have also reported similar findings in the porcine model, with western blot of stented porcine coronary arteries at 90 days demonstrating ongoing cellular proliferation despite elevation of p27kip1, a mediator of the antiproliferative effects of rapamycin. These data support the notion that polymers in DES might provoke chronic inflammation, which might be a potential driver of intimal formation, diminishing over a long term with decreased efficacy of these systems.
In contrast, pf-RES provoke little if any inflammatory reaction in animal models, but they also do not delay healing compared with BMS in the rabbit model. Although supporting the biocompatibility of this system, these data call into question the clinical antirestenotic efficacy reported by Byrne et al. (7). Previous nonpolymeric delivery of paclitaxel with a different system as reported by Park et al. (16) associated with less neointimal formation at 6 months, but this benefit was lost at 2 years especially in the high-dose paclitaxel arm. Heterogeneity of patient characteristics and lesions in the study by Bryne et al. complicates its interpretation, especially given that there were 2 follow-up time points. It is entirely possible that the characteristics of each group at early and late follow-up were different between stent arms, especially because patients who underwent revascularization were excluded for late follow-up analysis. In addition, the late restenosis rate is higher than has been reported by others, with composite in-segment binary restenosis being 17.3% for RES, 16.6% for pf-RES, and 21.8% for PES.
In summary, we believe these data support the notion that the vascular responses to pf-RES result in less delay in healing and diminished inflammatory responses compared with RES or PES but might be burdened with fewer efficacies. We can surmise that the data of Byrne et al. support the concept that, although DES delay intimal formation and healing, they do not halt it. As seen in animals, the 2 primary mechanisms responsible for driving the increase in intimal formation between 6 to 8 months and 2 years in humans is likely: 1) ongoing arterial repair that is accelerated in the face of decreasing drug levels; and 2) chronic inflammatory responses to polymer or excessive fibrin deposition, with both processes provoking cellular proliferation.
Practically speaking, these data also inform the way we should care for patients after DES placement. Whereas the surveillance period for symptoms of restenosis in patients receiving BMS is usually <6 months, clinicians need to be aware that polymeric DES have altered the temporal course of restenosis. In the study of Byrne et al. (7), of the 1,471 non-restenotic lesions at 6 to 8 months, 12% met the criteria for binary angiographic restenosis at 2 years, which is higher than that reported at 6 to 8 months, whereas pf-RES had higher restenosis at 6 to 8 months, but the restenosis did not increase at 2 years. These data highlight the need for continued surveillance in these patients well beyond that needed for patients receiving BMS and reminds us that, as we enlarge our understanding of DES technology, continued refinements in the care of these patients are needed.
 |
Footnotes
|
|---|
Dr. Virmani has received research support from 3F Therapeutics; Abbott Vascular; Amaranth Medical, Inc.; Apnex Medical; Ardian, Inc.; Atrium Medical Corporation; CardioDex, Ltd.; CardioKinetix, Inc.; CorAssist Cardiovascular, Ltd.; Cordis Corporation; Devax, Inc.; ev3; Gardia Medical, Ltd.; GlaxoSmithKline; HemCon; Lutonix, Inc.; Medtronic Vascular; Meril Life Sciences Pvt, Ltd.; Microvention, Inc.; Novartis Pharmaceuticals Corporation; NovoStent Corporation; Oregon Medical Laser Center; Prescient Medical, Inc.; Relisys Medical Devices Limited; Vascular Therapies, LLC; and Xtent, Inc. He has served as consultant to Medtronic AVE; Abbott Vascular; W.L. Gore; Volcano Therapeutics, Inc.; Prescient Medical; CardioMind, Inc.; Direct Flow; and Atrium Medical Corporation.
* Editorials published in JACC: Cardiovascular Interventions reflect the views of the authors and do not necessarily represent the views of JACC: Cardiovascular Interventions or the American College of Cardiology. 
* Reprint requests and correspondence: Dr. Renu Virmani, CVPath, Institute, Inc., 19 Firstfield Road, Gaithersburg, Maryland 20878 (Email: rvirmani{at}cvpath.org).
 |
REFERENCES
|
|---|
- Moses JW, Leon MB, Popma JJ, et al. Sirolimus-eluting stents versus standard stents in patients with stenosis in a native coronary artery N Engl J Med 2003;349:1315-1323.[CrossRef][Web of Science][Medline]
- Stone GW, Ellis SG, Cox DA, et al. A polymer-based, paclitaxel-eluting stent in patients with coronary artery disease N Engl J Med 2004;350:221-231.[CrossRef][Web of Science][Medline]
- Klugherz BD, Llanos G, Lieuallen W, et al. Twenty-eight-day efficacy and phamacokinetics of the sirolimus-eluting stent Coron Artery Dis 2002;13:183-188.[CrossRef][Web of Science][Medline]
- Carter AJ, Aggarwal M, Kopia GA, et al. Long-term effects of polymer-based, slow-release, sirolimus-eluting stents in a porcine coronary model Cardiovasc Res 2004;63:617-624.[Abstract/Free Full Text]
- Wilson GJ, Polovick JE, Huibregtse BA, Poff BC. Overlapping paclitaxel-eluting stents: long-term effects in a porcine coronary artery model Cardiovasc Res 2007;76:361-372.[Abstract/Free Full Text]
- Joner M, Finn AV, Farb A, et al. Pathology of drug-eluting stents in humans: delayed healing and late thrombotic risk J Am Coll Cardiol 2006;48:193-202.[Abstract/Free Full Text]
- Byrne RA, Iijima R, Mehilli J, et al. Durability of antirestenotic efficacy in drug-eluting stents with and without permanent polymer J Am Coll Cardiol Intv 2009;2:291-299.[Abstract/Free Full Text]
- Finn AV, Nakazawa G, Joner M, et al. Vascular responses to drug eluting stents: importance of delayed healing Arterioscler Thromb Vasc Biol 2007;27:1500-1510.[Abstract/Free Full Text]
- Virmani R, Kolodgie FD, Farb A, Lafont A. Drug eluting stents: are human and animal studies comparable? Heart 2003;89:133-138.[Abstract/Free Full Text]
- Wessely R, Hausleiter J, Michaelis C, et al. Inhibition of neointima formation by a novel drug-eluting stent system that allows for dose-adjustable, multiple, and on-site stent coating Arterioscler Thromb Vasc Biol 2005;25:748-753.[Abstract/Free Full Text]
- Stone G. 4 Year Clinical Results of TAXUS IV Presented at: 55th American College of Cardiology Annual Meeting, Atlanta, GA. 2006.
- Leon MB. 4 Year Outcomes in the SIRIUS Trial Presented at: Transcatheter Cardiovascular Therapeutics Annual Meeting. 2006Washington, DC.
- Kimura T, Yokoi H, Nakagawa Y, et al. Three-year follow-up after implantation of metallic coronary-artery stents N Engl J Med 1996;334:561-566.[CrossRef][Web of Science][Medline]
- Nakazawa G, Finn AV, Ladich E, et al. Drug-eluting stent safety: findings from preclinical studies Expert Rev Cardiovasc Ther 2008;6:1379-1391.[CrossRef][Medline]
- Finn AV, Kolodgie FD, Harnek J, et al. Differential response of delayed healing and persistent inflammation at sites of overlapping sirolimus- or paclitaxel-eluting stents Circulation 2005;112:270-278.[Abstract/Free Full Text]
- Park DW, Hong MK, Mintz GS, et al. Two-year follow-up of the quantitative angiographic and volumetric intravascular ultrasound analysis after nonpolymeric paclitaxel-eluting stent implantation: late "catch-up" phenomenon from ASPECT study J Am Coll Cardiol 2006;48:2432-2439.[Abstract/Free Full Text]
Related Article
-
Durability of Antirestenotic Efficacy in Drug-Eluting Stents With and Without Permanent Polymer
- Robert A. Byrne, Raisuke Iijima, Julinda Mehilli, Susanne Pinieck, Olga Bruskina, Albert Schömig, and Adnan Kastrati
J. Am. Coll. Cardiol. Intv. 2009 2: 291-299.
[Abstract]
[Full Text]
[PDF]
This article has been cited by other articles:

|
 |

|
 |
 
N. Foin, R. Torii, P. Mortier, M. De Beule, N. Viceconte, P. H. Chan, J. E. Davies, X. Y. Xu, R. Krams, and C. Di Mario
Kissing Balloon or Sequential Dilation of the Side Branch and Main Vessel for Provisional Stenting of Bifurcations: Lessons From Micro-Computed Tomography and Computational Simulations
J. Am. Coll. Cardiol. Intv.,
January 1, 2012;
5(1):
47 - 56.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. A. Byrne, A. Kastrati, K. Tiroch, S. Schulz, J. Pache, S. Pinieck, S. Massberg, M. Seyfarth, K.-L. Laugwitz, K. A. Birkmeier, et al.
2-Year Clinical and Angiographic Outcomes From a Randomized Trial of Polymer-Free Dual Drug-Eluting Stents Versus Polymer-Based Cypher and Endeavor, Drug-Eluting Stents
J. Am. Coll. Cardiol.,
June 8, 2010;
55(23):
2536 - 2543.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. B. Leon, D. E. Kandzari, E. L. Eisenstein, K. J. Anstrom, L. Mauri, D. E. Cutlip, E. Nikolsky, C. O'Shaughnessy, P. A. Overlie, A. J. Kirtane, et al.
Late Safety, Efficacy, and Cost-Effectiveness of a Zotarolimus-Eluting Stent Compared With a Paclitaxel-Eluting Stent in Patients With De Novo Coronary Lesions: 2-Year Follow-Up From the ENDEAVOR IV Trial (Randomized, Controlled Trial of the Medtronic Endeavor Drug [ABT-578] Eluting Coronary Stent System Versus the Taxus Paclitaxel-Eluting Coronary Stent System in De Novo Native Coronary Artery Lesions)
J. Am. Coll. Cardiol. Intv.,
December 1, 2009;
2(12):
1208 - 1218.
[Abstract]
[Full Text]
[PDF]
|
 |
|
|