A Randomized, Controlled, Multicenter Trial to Evaluate the Safety and Efficacy of Zotarolimus- Versus Paclitaxel-Eluting Stents in De Novo Occlusive Lesions in Coronary ArteriesThe ZoMaxx I Trial
Bernard Chevalier, MD*,
Carlo Di Mario, MD ,
Franz-Josef Neumann, MD ,
Flavio Ribichini, MD ,
Philip Urban, MD||,
Jeffrey J. Popma, MD¶,
Peter J. Fitzgerald, MD, PhD#,
Donald E. Cutlip, MD**,
David O. Williams, MD ,
John Ormiston, MD ,
Eberhard Grube, MD ,
Robert Whitbourn, MD||||,
Lewis B. Schwartz, MD¶¶,* for the ZoMaxx I Investigators
* Centre Cardiologique du Nord; Saint-Denis, France
Royal Brompton Hospital; London, England
Herz-Zentrum Bad Krozingen; Bad Krozingen, Germany
Università del Piemonte Orientale; Novara, Italy
|| La Tour Hospital; Geneva, Switzerland
¶ Brigham & Women's Hospital; Boston, Massachusetts
# Stanford University Medical Center; Palo Alto, California
** Harvard Clinical Research Institute; Boston, Massachusetts
 Rhode Island Hospital; Providence, Rhode Island
 Auckland City Hospital; Auckland, New Zealand
 Heart Center Siegburg; Siegburg, Germany
|||| St. Vincent's Hospital; Melbourne, Australia
¶¶ Abbott Laboratories; Abbott Park, Illinois
 |
Abstract
|
|---|
Objectives: A novel zotarolimus-eluting coronary stent system (ZoMaxx, Abbott Laboratories, Abbott Park, Illinois) was compared with a paclitaxel-eluting coronary stent (Taxus Express2) in a randomized trial of percutaneous intervention for de novo coronary artery stenosis. The primary end point was defined as noninferiority of in-segment late lumen loss after 9 months.
Background: The ZoMaxx stent system elutes 10 µg/mm zotarolimus using a phosphorylcholine polymer loaded onto a novel stainless steel stent platform containing a 0.0007-inch inner layer of tantalum.
Methods: Twenty-nine investigative sites in Europe, Australia, and New Zealand enrolled 401 patients, 396 of whom received a study stent.
Results: After 9 months, late lumen loss was significantly greater in the ZoMaxx group (in-stent 0.67 ± 0.57 mm vs. 0.45 ± 0.48 mm; p < 0.001; in-segment 0.43 ± 0.60 mm vs. 0.25 ± 0. 45 mm; p = 0.003), resulting in significantly higher rates of >50% angiographic restenosis (in-stent 12.9% vs. 5.7%; p = 0.03; in-segment 16.5% vs. 6.9%; p = 0.007). The upper bound of the 95% confidence interval on the difference in in-segment late lumen loss between the 2 treatment groups (0.27 mm) exceeded the 0.25 mm value pre-specified for noninferiority. There were no significant differences between ZoMaxx and Taxus-treated groups with respect to target lesion revascularization (8.0% vs. 4.1%; p = 0.14), major adverse cardiac events (12.6% vs. 9.6%; p = 0.43), or stent thrombosis (0.5% in both groups).
Conclusions: After 9 months, the ZoMaxx stent showed less neointimal inhibition than the Taxus stent, as shown by higher in-stent late loss and restenosis by qualitative coronary angiography.
Key Words: drug-eluting stent stent zotarolimus restenosis coronary artery disease
|
Abbreviations and Acronyms
| | CK = creatine kinase | | DES = drug-eluting stents | | IVUS = intravascular ultrasound | | MACE = major adverse cardiovascular events | | MLD = minimum lumen diameter | | PC = phosphorylcholine | | TLR = target lesion revascularization | | TVR = target vessel revascularization |
|
Drug-eluting stents (DES) are intravascular metal scaffolds that are coated with antiproliferative agents designed to treat critical occlusive lesions of the coronary arteries. Stents eluting either sirolimus, paclitaxel, everolimus, or zotarolimus have each been shown to effectively inhibit restenosis in large-scale clinical trials (1–4). First introduced in 2001, DES have been widely applied, with over 6 million patients treated worldwide (5).
The clinical trial reported herein was designed to compare the novel ZoMaxx zotarolimus-eluting stent (Abbott Laboratories, Abbott Park, Illinois) with the Taxus Express2 paclitaxel-eluting stent (Boston Scientific Corporation, Natick, Massachusetts). The ZoMaxx stent uses the TriMaxx (Abbott Laboratories) stainless steel–tantalum stent platform to deliver zotarolimus 10 µg/mm via a well-characterized polymer system based on phosphorylcholine (PC) (6). The objective of this randomized, prospective, multicenter trial was to show the safety and efficacy of the ZoMaxx stent system as compared with the Taxus Express2 stent system for patients with single de novo lesions in native coronary arteries using clinical, angiographic, and intravascular ultrasonic methods.
 |
Methods
|
|---|
Study design and end points.
The ZoMaxx I trial was a randomized, prospective, multicenter clinical trial conducted in accordance with the International Conference on Harmonization guidelines–Good Clinical Practices, Declaration of Helsinki, International Organization for Standardization 14155-1, International Organization for Standardization 14155-2, and Ethics Committee requirements. All patients gave written informed consent for participation.
Patients were considered eligible for inclusion if they complained of stable or unstable angina and/or had objective evidence of myocardial ischemia with angiographically proven single >50% lesions of 10 to 30 mm in length in 2.5- to 3.5-mm native coronary arteries. The major clinical exclusion criteria were acute myocardial infarction within the past 72 h, impaired left ventricular function with ejection fraction <30%, or lesions located within the left main coronary artery or within 2.0 mm of their ostia.
Secondary end points included device success (achievement of a final residual in-stent diameter stenosis of <30% using the assigned device only), lesion success (<30% residual stenosis using any percutaneous method), procedure success (lesion success without the occurrence of major adverse cardiac events [MACE]), angiographic rates of binary restenosis ( 50% diameter stenosis) after 9 months, neointimal hyperplasia volume after 9 months as measured by intravascular ultrasound (IVUS), and the 9-month incidences of ischemia-driven target lesion revascularization (TLR), ischemia-driven target vessel revascularization (TVR), and MACE (composite end point of non–Q-wave myocardial infarction [MI], Q-wave MI, TVR, and cardiac death). The first 250 randomized subjects with IVUS-eligible lesions were required to have IVUS evaluation during the procedure and at 9-month follow-up. A Q-wave MI was defined as the development of new pathological Q-waves in 2 or more contiguous leads with post-procedure creatine kinase (CK) or CK-MB levels elevated above normal. Non–Q-wave MI was defined as elevation of post-procedure CK levels to >2.0 times normal with elevated CK-MB in the absence of new pathological Q waves (World Health Organization definition). Acute luminal gain was defined as the difference between the minimum lumen diameter (MLD) at the completion of the stenting procedure and at baseline. Data are expressed as the mean ± SD for continuous variables and as frequencies for categorical variables (SAS Statistical Analysis Software for Windows, version 8.2, SAS Institute Inc., Cary, North Carolina).
Stent system.
The ZoMaxx stent was designed to address the need for thin strut width and low profile, while maintaining radial strength and adequate visibility on fluoroscopy. The stent metal is a trilayer composite having 2 outer layers of 316L stainless steel and an inner layer of tantalum (Fig. 1) (7). The high atomic number of the 18-µm inner tantalum layer affords optimal radiopacity of the thin stent struts. The result is a DES with a strut thickness of only 0.0029 inches (74 µm), an important metric for minimizing arterial injury and restenosis (8–10).
The ZoMaxx stent elutes 10 µg/mm zotarolimus via a biocompatible PC polymer. Zotarolimus (Fig. 2) was specifically developed for use on intravascular stents and, like sirolimus, reversibly binds to FKBP-12, the cytosolic receptor of FK506 (11,12). Using this mechanism, zotarolimus inhibits the activation and proliferation of a variety mammalian cells at very low concentrations. Its potency for inhibition of human lymphocytes in vitro has been shown, as well as the reduction of inflammation in animal models of arthritis and encephalomyelitis (13). In cultured vascular cells, zotarolimus inhibits proliferation of canine and human smooth muscle and endothelial cells with IC50 in the low nanomolar range (11,12,14). It has minimal effects on cellular migration in vitro, theoretically allowing re-endothelialization to proceed normally in the presence of the drug.
The phosphorylcholine polymer drug carrier on the ZoMaxx stent, known simply as PC-1036 or PC, is composed of the polymers 2-methacryloyloxyethyl phosphorylcholine (MPC), lauryl methacrylate (LMA), hydroxypropyl methacrylate (HPMA), and trimethoxysilylpropyl methacrylate (TSMA) in the molar ratios of MPC23, LMA47, HPMA25, and TSMA5 (6,15). It has been experimentally shown to have several properties improving blood–biomaterial compatibility, including minimal induction of fibrinogen absorption, platelet activation, platelet and erythrocyte adherence (15–17), inflammation, and neointimal hyperplasia (6,17–20). Phosphorylcholine has an extensive record of widespread clinical use and safety (6,21–23). It has also recently been shown to have considerable potential as a vehicle for drug elution, especially using highly lipophilic agents. It is formulated on the ZoMaxx stent to provide a measured rate of elution so that, in experimental animals, about 60% of the total zotarolimus dose is released during the first week, an additional 20% during the second week, and the remaining 20% over the next 2 weeks (24).
Treatment protocol.
After patient eligibility was established and written consent obtained, the lesions were approached according to standard institutional interventional techniques. Patients received oral antiplatelet therapy with aspirin (100 mg/day) and clopidogrel (75 mg/day) starting before the procedure and continuing for 6 months. After percutaneous access, heparin was administered to maintain an activated clotting time 250 s (or 200 s if glycoprotein IIb/IIIa antagonism was used). Diagnostic coronary angiography was performed in matched orthogonal views after nitroglycerin coronary injection (50 to 200 µg). After a 0.014-inch wire crossing of the target lesion, randomization (1:1 ZoMaxx vs. Taxus) was performed via an interactive telephone system.
Balloon pre-dilation of the target lesion was mandatory and was performed according to standard techniques ensuring that the length of the balloon was no greater than the intended stent length. Direct stenting was prohibited in this study. ZoMaxx stents were available in diameters of 2.5, 3.0, and 3.5 mm with lengths of 8, 18, 23, 28 mm (2.5-mm diameter only), and 33 mm (3.0- and 3.5-mm diameters only). Taxus stents were available in diameters of 2.5, 3.0, and 3.5 mm with lengths of 8, 12, 16, 20, 24, 28 mm (3.0- and 3.5-mm diameters only) and 32 mm (3.0- and 3.5-mm diameters only). Only 1 study stent was to be used per patient; however, additional stents could be implanted at the operator's discretion in the event of edge dissection or incomplete coverage.
Intravascular ultrasound images were acquired by motorized pullback at a constant speed of 0.5 mm/s. Baseline, post-procedure, and 9-month follow-up coronary cineangiographic images (Medis CMS, Leiden, the Netherlands) and IVUS tapes (TapeMeasure, Indec Systems, Inc., Mountain View, California) were analyzed using independent core laboratories (Brigham and Women's Hospital, Boston, Massachusetts, and Stanford University Medical Center, Palo Alto, California, respectively). Imaging studies performed within 284 days (9 months + 2 weeks) were included in the analysis.
Monitoring and statistical analysis.
The study was monitored by independent contract research organizations (Hesperion AG, Allschwil, Switzerland, and Clinimetrics Research Associates, Inc., San Jose, California) and data were coordinated and analyzed by the Harvard Clinical Research Institute (Boston, Massachusetts). All MACE were reviewed and adjudicated by an independent Clinical Events Committee whose members were unaware of treatment allocation. An independent Data and Safety Monitoring Board periodically reviewed blinded safety data.
The trial was designed to show noninferiority of in-segment late loss after 9 months (expected difference in means, µZoMaxx Stent – µTaxus Stent = 0), with a noninferiority margin of 0.25 mm and a standard deviation of 0.4 mm. Late luminal loss was defined as the difference between the MLD immediately after stenting and at follow-up. Because late loss values <0.6 mm for stented arteries between 2.5 and 3.5 mm in diameter have little clinical consequence, and the Taxus stent can be expected to generate 0.23 ± 0.44 mm in-segment late loss in this lesion cohort (2), a noninferiority margin of 0.25 mm was considered clinically and statistically appropriate. The trial was 99% powered to detect noninferiority with a 1-sided p value of 0.05.
Other treatment group comparisons were performed using the 2-sample t test for continuous variables, Fisher exact test for dichotomous variables, and Cochran-Mantel-Haenszel (Modified Ridit scores) for ordinal variables with more than 2 categories. Univariate and multivariate logistic regression was performed on the binary TVR results to understand the predictive value of several covariates. For the multivariate logistic regression, predictors were chosen by stepwise linear regression using an entry criteria of 0.2 and a stay criteria of 0.1.
 |
Results
|
|---|
Four-hundred and one patients were enrolled sequentially in the study from 29 clinical sites in Europe, Australia, and New Zealand (Appendix). The first patient was enrolled on September 14, 2004, and the final patient was enrolled on July 18, 2005. Five patients (4 randomized to ZoMaxx and 1 randomized to Taxus) were subsequently deregistered after randomization and did not receive a study stent (3 were deemed to be ineligible after randomization, 1 sustained a complication before stent insertion, and 1 withdrew consent before stent insertion). This left a total of 396 patients for analysis (199 ZoMaxx and 197 Taxus).
The clinical and lesional demographics of the 2 patient cohorts are given in Table 1. The groups were fairly well matched demographically, as there were similar frequencies of diabetes (ZoMaxx 22% vs. Taxus 26%) and unstable angina (ZoMaxx 26% vs. Taxus 24%). However, there was statistically significantly more intervention in the right coronary artery in the Taxus group versus the ZoMaxx group (41% vs. 28%; p = 0.008). Furthermore, 8 lesions in the ZoMaxx group were ostial in location (within 2 mm of their origin); there were no ostial lesions in the Taxus group (4.0% vs. 0%; p = 0.0007).
Post-procedure metrics are given in Table 2. Lesion and device success were 99% for both stents (p = NS). There were no differences with respect to post-procedure in-stent or in-segment MLD, percent diameter stenosis, or acute gain between the 2 groups.
Angiographic results are given in Table 3. Nine-month in-segment late lumen loss (primary end point) was 0.43 ± 0.60 mm and 0.25 ± 0.45 mm in the ZoMaxx and Taxus angiographic cohorts, respectively; the observed difference in the means was 0.18 mm. The distributions of in-segment late lumen loss in the ZoMaxx and Taxus groups are shown in Figure 3. Because the upper bound of a 95% confidence interval on the difference in the means (0.27 mm) was larger than the pre-specified noninferiority limit (0.25 mm), the primary angiographic end point of noninferiority of in-segment late lumen loss was not met. Consistent with the in-segment late loss results, in-segment binary restenosis was greater in the ZoMaxx group than in the Taxus group (16.5% vs. 6.9%; p = 0.007).
Similarly, in-stent late lumen loss was found to be statistically significantly greater after ZoMaxx stenting as compared with Taxus stenting (0.67 ± 0.57 mm vs. 0.45 ± 0. 48 mm; p < 0.0001), resulting in a higher frequency of in-stent restenosis in the ZoMaxx group (12.9% vs. 5.7%; p = 0.025). Even using nonparametric analysis, more appropriate for non-normal distributions such as late loss after implantation of DES (25,26), the difference in the medians remained statistically significant (median in-stent late loss ZoMaxx 0.58 mm vs. Taxus 0.41 mm; p < 0.05 using the Kruskal-Wallis test).
Clinical results at 9 months encompassing all follow-up angiography performed up to 284 days are given in Table 4; 9-month clinical follow-up was available in 96% of patients (382 of 396). There were no significant differences between treatment groups for any clinical end point. The rate of TLR was nearly double in the ZoMaxx group as compared with the Taxus group (8.0% vs. 4.1%), although the difference did not reach statistical significance (p = NS). There were no differences in the incidence of stent thrombosis whether protocol-defined (0.5% in both groups) or retrospectively applying the definitions suggested by the Academic Research Consortium (27) (1.0% in both groups).
View this table:
[in this window]
[in a new window]
|
Table 4 9-Month Clinical Results in the ZoMaxx I Trial (Includes All Follow-Up Angiograms Performed Through 284 Days)
|
|
The IVUS results are given in Table 5. Neointimal volume obstruction by IVUS was statistically significantly greater after ZoMaxx stenting (14.6 ± 7.9% vs. 11.2 ± 9.6 %; p < 0.018). The incidence of late acquired malapposition was slightly less after ZoMaxx stenting, although the difference did not reach statistical significance (0% vs. 3%; p = NS).
To evaluate the possible predictive value of covariates in the ZoMaxx I study, both univariate and multivariable logistic regression analyses were performed on the entire patient cohort to identify significant risk factors for the need for follow-up TVR. The results, shown in Table 6, identify only ostial lesion location (p = 0.002) and the presence of diabetes (p = 0.003) as statistically significant predictors of the need for TVR after 9 months.
 |
Discussion
|
|---|
Although a variety of antiproliferative agents have been suggested as putative inhibitors of stent-induced restenosis, only sirolimus (1), paclitaxel (2), everolimus (3), and zotarolimus (28) have been proven safe and effective in large-scale, multicenter, randomized clinical trials. One compound, zotarolimus (formerly known as ABT-578, Abbott Laboratories) (Fig. 2), has been specifically developed for use in DES having no other systemic formulation or indication. Intravascular stents that elute zotarolimus have been shown to be effective in inhibiting in-stent restenosis both in experimental animal models (29) and in patients undergoing percutaneous coronary intervention (4,28).
To date, 4 zotarolimus-eluting devices have been tested clinically: the Prefer (Abbott Laboratories) (30), Endeavor (Medtronic, Minneapolis, Minnesota), (4) ZoMaxx (31), and Resolute (Medtronic) (32) stents. The most extensively studied is the Endeavor stent, which also contains 10 µg/ml zotarolimus in a PC-based formulation that, in experimental animals, elutes approximately 95% of its drug load over about 15 days (4). The Endeavor II pivotal trial compared the Endeavor stent to the bare metal Driver stent in 1,200 patients, and the results showed highly statistically significant reductions in late lumen loss (in-stent 1.03 ± 0.58 mm vs. 0.61 ± 0.46 mm; p < 0.0001), angiographic binary restenosis (in-stent 33.5% vs. 9.4%; p < 0.0001), target vessel failure (15.1% vs. 7.9%; p = 0.0001) and MACE (14.4% vs. 7.3%; p = 0.0001) (28). Based on these and other clinical trial results (4,33,34), the Endeavor stent is now approved for clinical use worldwide.
The ZoMaxx stent was first tested clinically in the ZoMaxx IVUS trial, which enrolled 40 patients with symptomatic ischemic coronary occlusive disease at the Instituto Dante Pazzanese de Cardiologia in Saõ Paulo, Brazil (31). The lesion, procedure, and device-deployment success rates were all 100% (40 of 40), and there was no MACE during the 4-month study. Follow-up angiography at 4 months showed in-stent and in-segment late lumen losses of 0.20 ± 0.35 mm and 0.17 ± 0.35 mm, respectively, with IVUS examinations showing a mean of 6.5 ± 6.2% neointimal volume obstruction. There were no instances of late acquired stent incomplete apposition or stent thrombosis. In comparison with a similar clinical trial using the nondrug TriMaxx stent studied angiographically at 6 months, the ZoMaxx stent significantly reduced in-stent restenosis from 25% to 2.7% (35).
These initial observations were extended through the ZoMaxx I clinical trial reported herein. The ZoMaxx I trial was designed to study a larger and more complex patient cohort with longer duration of follow-up in multinational geographies and to concurrently compare the outcome of ZoMaxx stent implantation with patients treated with the Taxus paclitaxel-eluting stent.
The ZoMaxx and Taxus patient groups in the ZoMaxx I trial were generally well-matched in both clinical and lesion characteristics. The exception was the preponderance of ostial lesions within the ZoMaxx cohort (n = 8) as compared with the Taxus cohort (n = 0). This occurrence was random and was uncovered as a result of retrospective core laboratory analysis of baseline angiograms, wherein lesions believed to be acceptable candidates for entry into the study were subsequently found to be within 2 mm of the artery's origin. It is known that percutaneous treatment of ostial occlusive lesions carries a substantially higher risk of restenosis (36–39), TVR (37), and 1-year mortality (40), and this is why patients with lesions that are located near one of the coronary ostia have been specifically excluded from pivotal clinical trials of DES (1,2,28). Indeed, the presence of an ostial lesion was the most significant multivariate predictor of TVR in the 386-patient cohort (p = 0.002), even slightly stronger than the presence of diabetes mellitus (p = 0.003).
The inclusion of patients with ostial lesions in the ZoMaxx group notwithstanding, there were no differences between the ZoMaxx and Taxus stent in device deployment or safety as shown by the high rates of device success (99% in both groups), a low stent thrombosis rate, and the absence of late stent thrombosis. There were no significant differences between treatment groups for any clinical metric that was evaluated (Table 4), although the study was not specifically powered to detect differences in infrequent events.
The most important finding in this clinical study was the statistically significant difference in late lumen loss between the 2 angiographic cohorts. The frequencies of both angiographic (87%) and IVUS (59%) follow-up are among the highest reported in any clinical trial of DES. After 9 months, patients treated with the ZoMaxx stent showed a mean in-segment late lumen loss of 0.43 ± 0.60 mm, whereas patients treated with the Taxus stent showed only 0.25 ± 0.45 mm of luminal loss (p < 0.001). Because the upper bound of a 95% confidence interval on the difference between treatment groups for in-segment late loss was 0.27 mm (larger than the pre-specified 0.25 mm), the primary angiographic end point of noninferiority of in-segment late lumen loss was not met.
Given the differences in late lumen loss and the well-described curvilinear relationship of late loss to restenosis (25), it is not surprising that the ZoMaxx group was found to have statistically significantly higher rates of in-stent restenosis (12.9% vs. 5.7%; p = 0.025), as well as a trend toward increased ischemia-driven TLR as compared with Taxus (8.0% vs. 4.1%; p = NS). The specific mechanisms underlying these findings are speculative. The 2 stent systems differ in all critical design elements, including their stent platforms, drugs, pharmacological mechanisms of action, polymers, and formulations. The nondrug bare metal stent platforms seem to yield roughly comparable clinical and angiographic results, bearing in mind the relatively small patient cohorts subjected to rigorous angiographic follow-up (2,7,35). Both drugs have been shown to effectively inhibit mammalian cell proliferation in vitro and in animal models (29,41), and both have yielded variable clinical and angiographic results depending on their formulations, intervals of angiographic follow-up, and specific populations under study (range of in-stent late loss for paclitaxel-eluting stents: 0.30 to 0.81 mm [2,42–45]; zotarolimus-eluting stents: 0.12 to 0.67 mm [28,32,33,46]). It is noteworthy that the 9-month in-stent late lumen loss of patients treated with the ZoMaxx stent in this study (0.67 ± 0.57 mm) is strikingly similar to 8-month late lumen loss of patients treated with the Endeavor stent in each of Endeavor II (0.62 ± 0.46 mm) (28), Endeavor III (0.60 ± 0.48 mm) (33) and Endeavor IV (0.67 ± 0.49 mm) clinical trials (47). Thus, the prolonged release rate of ZoMaxx (24) as compared with Endeavor showed in nonconcurrent animal testing (4,24) had no apparent effect on results in humans. Interestingly, a zotarolimus-eluting stent using a different polymer formulation and having an even longer elution rate than ZoMaxx has recently been developed (Endeavor Resolute, Medtronic). Preliminary angiographic results in 30 patients suggest enhanced inhibition of neointimal hyperplasia using this formulation with a mean in-stent late loss of 0.12 ± 0.26 mm after 4 months (32). It can only be concluded that the efficacy of a given DES continues to be difficult to predict empirically and that long-term comparative clinical testing of each new formulation is required before its widespread application.
 |
Conclusions
|
|---|
The ZoMaxx Coronary Stent can be safely implanted for the treatment of de novo coronary artery stenosis, as evidenced by the high rate of device implantation success (99%) and the low rates of subacute (0.5%) and late (0%) stent thrombosis. After 9 months, the ZoMaxx stent showed less neointimal inhibition than the Taxus stent, as demonstrated by higher in-stent late loss and restenosis by QCA and neointimal volume obstruction by IVUS.
 |
Appendix
|
|---|
For a complete list of investigators and institutions, please see the online version of this article.
 |
Footnotes
|
|---|
Dr. Chevalier has received research support from and is a consultant for Abbott Laboratories; Dr. DiMario is on the Speaker's Bureau and has received honoraria and/or is a consultant for Abbott Laboratories and Boston Scientific; Dr. Urban has received research support from, is on the Speaker's Bureau, and/or is a consultant for Biosenors, Johnson & Johnson, Medtronic, and Terumo; Dr. Popma has received research support from, is on the Speaker's Bureau of, and/or is a consultant for Abbott Laboratories, Boston Scientific, eV3, Johnson & Johnson, Eli Lilly and Company, Medtronic, and Sanofi; Dr. Fitzgerald has received research support from and is a consultant for Abbott Laboratories; Dr. Williams is a consultant for Abbott Laboratories; Dr. Ormiston has received research support or honoraria from and/or is a consultant for Abbott Laboratories and Boston Scientific; Dr. Schwartz is a full-time employee of Abbott Laboratories.
* Reprint requests and correspondence: Dr. Lewis B. Schwartz, Abbott Laboratories, 200 Abbott Park Road, AP52-2, Abbott Park, Illinois 60064-6215 (Email: lewis.schwartz{at}abbott.com).
Manuscript received June 30, 2008;
revised manuscript received July 21, 2008,
accepted August 20, 2008.
 |
REFERENCES
|
|---|
- Moses JW, Leon MB, Popma JJ, et al. Sirolimus-eluting stents versus standard stents in patients with stenosis in a native coronary artery N Engl J Med 2003;349:1315-1323.[Abstract/Free Full Text]
- Stone GW, Ellis SG, Cox DA, et al. A polymer-based paclitaxel-eluting stent in patients with coronary artery disease N Engl J Med 2004;350:221-231.[Abstract/Free Full Text]
- Serruys PW, Ruygrok P, Neuzner J, et al. A randomised comparison of an everolimus-eluting coronary stent with a paclitaxel-eluting coronary stent: the SPIRIT II trial Eurointervention 2006;2:286-294.
- Kandzari DE, Leon MB. Overview of pharmacology and clinical trials program with the zotarolimus-eluting Endeavor stent J Interv Cardiol 2006;19:405-413.[CrossRef][Medline]
- Benesh P. Market for Drug-Eluting Stents Grows CrowdedInvestor's Business Daily, November 2 2007.
- Lewis AL, Stratford PW. Phosphorylcholine-coated stents J Long Term Eff Med Implants 2002;12:231-250.[Medline]
- Abizaid A, Popma JJ, Tanajura LF, et al. Clinical and angiographic results of percutaneous coronary revascularization using a trilayer stainless steel-tantalum-stainless steel phosphorylcholine-coated stent: the TriMaxx trial Catheter Cardiovasc Interv 2007;70:914-919.[CrossRef][Web of Science][Medline]
- Hausleiter J, Kastrati A, Mehilli J, et al. Impact of lesion complexity on the capacity of a trial to detect differences in stent performance: results from the ISAR-STEREO trial Am Heart J 2003;146:882-886.[CrossRef][Web of Science][Medline]
- Pache J, Kastrati A, Mehilli J, et al. Intracoronary stenting and angiographic results: strut thickness effect on restenosis outcome (ISAR-STEREO-2) trial J Am Coll Cardiol 2003;41:1283-1288.[Abstract/Free Full Text]
- Kastrati A, Mehilli J, Dirschinger J, et al. Intracoronary stenting and angiographic results: strut thickness effect on restenosis outcome (ISAR-STEREO) trial Circulation 2001;103:2816-2821.[Abstract/Free Full Text]
- Chen Y-W, Smith ML, Sheets M, et al. Zotarolimus, a novel sirolimus analogue with potent anti-proliferative activity on coronary smooth muscle cells and reduced potential for systemic immunosuppression J Cardiovasc Pharmacol 2007;49:228-235.[CrossRef][Web of Science][Medline]
- Burke SE, Kuntz RE, Schwartz LB. Zotarolimus (ABT-578) eluting stents Adv Drug Deliv Rev 2006;58:437-446.[CrossRef][Web of Science][Medline]
- Wagner R, Mollison KW, Liu L, et al. Rapamycin analogs with reduced systemic exposure Bioorg Med Chem Lett 2005;15:5340-5343.[CrossRef][Medline]
- Xue L, Sharma R, Cochran K, et al. The effect of the rapamycin derivative ABT-578 on canine smooth muscle cell and endothelial cell proliferation Preclinica 2004;2:451-455.
- Lewis AL, Cumming ZL, Goreish HH, Kirkwood LC, Tolhurst LA, Stratford PW. Crosslinkable coatings from phosphorylcholine-based polymers Biomaterials 2001;22:99-111.[CrossRef][Web of Science][Medline]
- Campbell EJ, O'Byrne V, Stratford PW, et al. Biocompatible surfaces using methacryloylphosphorylcholine laurylmethacrylate copolymers ASAIO J 1994;40:853-857.[CrossRef]
- Lewis AL, Tolhurst LA, Stratford PW. Analysis of phosphorylcholine-based polymer coating on a coronary stent pre- and post-implantation Biomaterials 2002;23:1697-1706.[CrossRef][Web of Science][Medline]
- Malik N, Gunn J, Shepherd L, Crossman DC, Cumberland DC, Holt CM. Phosphorylcholine-coated stents in porcine coronary arteries: in vivo assessment of biocompatibility J Invasive Cardiol 2001;13:193-201.[Medline]
- Kuiper KK, Robinson KA, Chronos NAF, Cui J, Palmer SJ, Nordrehaug JE. Phosphorylcholine-coated stents in rabbit iliac and porcine coronary arteries Scand Cardiovasc J 1998;32:261-268.[CrossRef][Web of Science][Medline]
- Whelan DM, van der Giessen WJ, Krabbendam SC, et al. Biocompatibility of phosphorylcholine coated stents in normal porcine coronary arteries Heart 2000;83:338-345.[Abstract/Free Full Text]
- Zheng H, Barragan P, Corcos T, et al. Clinical experience with a new biocompatible phosphorylcholine-coated coronary stent J Invasive Cardiol 1999;11:608-614.[Web of Science][Medline]
- Galli M, Bartorelli A, Bedogni F, et al. Italian BiodivYsio Open Registry (BiodivYsio PC-coated stent): study of clinical outcomes of the implant of a PC-coated coronary stent J Invasive Cardiol 2000;12:452-458.[Web of Science][Medline]
- Boland JL, Corbeij HAM, van der Giessen WJ, et al. Multicentre evaluation of the phosphorylcholine-coated BiodivYsio stent in short de novo coronary lesions: the SOPHOS study Int J Cardiovasc Intervent 2000;3:215-225.[CrossRef][Medline]
- DuVall M, Ji Q, Clifford A, et al. ABT-578 elution profile and arterial penetration using the ZoMaxx drug-eluting stent (abstr) Am J Cardiol 2004;94:223E.
- Ellis SG, Popma JJ, Lasala JM, et al. Relationship between angiographic late loss and target lesion revascularization after coronary stent implantation: analysis from the TAXUS-IV Trial J Am Coll Cardiol 2005;45:1193-1200.[Abstract/Free Full Text]
- Kereiakes DJ, Kuntz RE, Mauri L, Krucoff MW. Surrogates, substudies, and real clinical end points in trials of drug-eluting stents J Am Coll Cardiol 2005;45:1206-1212.[Free Full Text]
- Cutlip DE, Windecker S, Mehran R, et al. Clinical end points in coronary stent trials: a case for standardized definitions Circulation 2007;115:2344-2351.[Abstract/Free Full Text]
- Fajadet J, Wijns W, Laarman G-J, et al. Randomized, double-blind, multicenter study of the Endeavor zotarolimus-eluting phosphorylcholine-encapsulated stent for treatment of native coronary artery lesions: clinical and angiographic results of the ENDEAVOR II trial Circulation 2006;114:798-806.[Abstract/Free Full Text]
- Garcia-Touchard A, Burke SE, Toner JT, Cromack K, Schwartz RS. Zotarolimus-eluting stents reduce experimental coronary artery neointimal hyperplasia after 4 weeks Eur Heart J 2006;27:988-993.[Abstract/Free Full Text]
- Sakurai R, Hongo Y, Ormiston J, et al. First human experience with the ABT-578 eluting phosphorylcholine polymer stent: a serial volumetric intravascular ultrasound analysis (abstr)Abstract presented at: American College of Cardiology; March 8 2004New Orleans, Louisiana.
- Abizaid A, Lansky AL, Fitzgerald PJ, et al. Percutaneous coronary revascularization using a trilayer phosphorylcholine-coated zotarolimus-eluting stent Am J Cardiol 2007;99:1403-1408.[CrossRef][Web of Science][Medline]
- Meredith IT, Worthly S, Whitbourn R, et al. The next-generation Endeavor Resolute stent: 4-month clinical and angiographic results from the Endeavor Resolute first-in-man trial Eurointervention 2007;3:50-53.
- Kandzari DE, Leon MB, Popma JJ, et al. Comparison of zotarolimus-eluting and sirolimus-eluting stents in patients with native coronary artery disease: a randomized controlled trial J Am Coll Cardiol 2006;48:2440-2447.[Abstract/Free Full Text]
- Jain AK, Meredith IT, Lotan C, Rothman MT, Pateraki S. Real-world safety and efficacy of the endeavor zotarolimus-eluting stent: early data from the E-Five Registry Am J Cardiol 2007;100:77M-83M.[Web of Science][Medline]
- Lasave LI, Costa Jr JDR, Abizaid AA, et al. A three-dimensional intravascular ultrasound comparison between the new zotarolimus-eluting stent (ZoMaxx) and the non-drug eluted TriMaxx stent J Invasive Cardiol 2007;19:303-308.[Medline]
- Satler L. Aorto-ostial disease and aorto-ostial in-stent restenosis: poorly recognized but very complex lesion subsets Catheter Cardiovasc Interv 2002;56:220-221.[CrossRef][Web of Science][Medline]
- Singh M, Gersh BJ, McClelland RL, et al. Predictive factors for ischemic target vessel revascularization in the prevention of restenosis with Tranilast and its outcomes (PRESTO) Trial J Am Coll Cardiol 2005;45:198-203.[Abstract/Free Full Text]
- Huang HL, Hsieh IC, Chang SH, et al. Acute and long-term outcomes of intracoronary stenting in aorto-ostial, left anterior descending artery-ostial and nonostial stenoses Int J Cardiol 2005;101:391-397.[CrossRef][Web of Science][Medline]
- Lemos PA, Hoye A, Goedhart D, et al. Clinical, angiographic, and procedural predictors of angiographic restenosis after sirolimus-eluting stent implantation in complex patients: an evaluation from the Rapamycin-Eluting Stent Evaluated At Rotterdam Cardiology Hospital (RESEARCH) study Circulation 2004;109:1366-1370.[Abstract/Free Full Text]
- Mavromatis K, Ghazzal Z, Veledar E, et al. Comparison of outcomes of percutaneous coronary intervention of ostial versus nonostial narrowing of the major epicardial coronary arteries Am J Cardiol 2004;94:583-587.[CrossRef][Web of Science][Medline]
- Heldman AW, Cheng L, Jenkins GM, et al. Paclitaxel stent coating inhibits neointimal hyperplasia at 4 weeks in a porcine model of coronary restenosis Circulation 2001;103:2289-2295.[Abstract/Free Full Text]
- Colombo A, Drzewiecki J, Banning A, et al. Randomized study to assess the effectiveness of slow- and moderate-release polymer-based paclitaxel-eluting stents for coronary artery lesions Circulation 2003;108:788-794.[Abstract/Free Full Text]
- Park SJ, Shim WH, Ho DS, et al. A paclitaxel-eluting stent for the prevention of coronary restenosis N Engl J Med 2003;348:1537-1545.[Abstract/Free Full Text]
- Lansky AJ, Costa RA, Mintz GS, et al. Non–polymer-based paclitaxel-coated coronary stents for the treatment of patients with de novo coronary lesions: angiographic follow-up of the DELIVER Clinical Trial Circulation 2004;109:1948-1954.[Abstract/Free Full Text]
- Stone GW, Ellis SG, Cannon L, et al. Comparison of a polymer-based paclitaxel-eluting stent with a bare metal stent in patients with complex coronary artery disease: a randomized controlled trial JAMA 2005;294:1215-1223.[Abstract/Free Full Text]
- Meredith IT, Ormiston J, Whitbourn R, et al. First-in-human study of the Endeavor ABT-578-eluting phosphorylcholine-encapsulated stent system in de novo native coronary artery lesions: Endeavor I Trial Eurointervention 2005;2:157-164.
- Leon MB. Endeavor IV: a randomized comparison of a zotarolimus-eluting stent and a paclitaxel-eluting stent in patients with coronary artery diseaseTranscatheter Cardiovascular Therapeutics. October 22 2007Washington, DC.
Related Article
-
The Process of Bringing New Drug-Eluting Stents to Market: Will They See the Light of Day?
- David R. Holmes, Jr and Manesh Patel
J. Am. Coll. Cardiol. Intv. 2008 1: 533-534.
[Full Text]
[PDF]
This article has been cited by other articles:

|
 |

|
 |
 
S. R. Dixon, C. L. Grines, and W. W. O'Neill
The year in interventional cardiology.
J. Am. Coll. Cardiol.,
June 2, 2009;
53(22):
2080 - 2097.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
D. R. Holmes Jr and M. Patel
The Process of Bringing New Drug-Eluting Stents to Market: Will They See the Light of Day?
J. Am. Coll. Cardiol. Intv.,
October 1, 2008;
1(5):
533 - 534.
[Full Text]
[PDF]
|
 |
|
|