J Am Coll Cardiol Intv, 2008; 1:22-31, doi:10.1016/j.jcin.2007.10.005
© 2008 by the American College of Cardiology Foundation
This Article
Right arrow Abstract Freely available
Right arrow Figures Only
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Related articles in JACC Interventions
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Romagnoli, E.
Right arrow Articles by Colombo, A.
Right arrow Search for Related Content
PubMed
Right arrow Articles by Romagnoli, E.
Right arrow Articles by Colombo, A.

Viewpoint

Drug-Eluting Stenting

The Case for Post-Dilation

Enrico Romagnoli, MD, PhD*, Giuseppe M. Sangiorgi, MD*, John Cosgrave, MD*, Edouard Guillet{dagger}, Antonio Colombo, MD*,*

* Emo Centro Cuore Columbus, Milan, Italy
{dagger} Medtronic Corporation, Tolochenaz, Switzerland.


    Abstract
 Top
 Abstract
 Diagnosis and Magnitude of...
 Determinants of Suboptimal Stent...
 High-Pressure Stent Deployment...
 Compliant Versus Noncompliant...
 Benefits of a Noncompliant...
 Conclusions
 REFERENCES
 
In clinical practice, adequate stent deployment has an important effect on immediate and long-term results after percutaneous coronary interventions. In particular, suboptimal or incomplete stent expansion is associated with increased restenosis and target vessel revascularization rates and, especially with drug-eluting stents (DES), might also predispose to stent thrombosis. Notwithstanding the significant improvement in technique and materials in the last decade, adjunctive high-pressure balloon dilation is still necessary to improve the minimum stent area and the uniform volumetric stent expansion in a majority of the cases. Indeed, in the published reports, the incidence of incomplete stent deployment ranges from 20% to 30% of cases, but it is significantly higher in trials in which stent expansion was assessed by intravascular ultrasound. Although there are not enough randomized studies about this topic, data from published reports continue to support the use of proper post-dilation in the majority of patients undergoing both bare-metal stent and DES implantation. This review will summarize the different anatomical, clinical, and device-related variables for increased risk of suboptimal stent delivery, highlighting the importance of adequate high-pressure post-dilation to obtain optimal stent expansion to positively affect stent thrombosis and restenosis.

Abbreviations and Acronyms
  BMS = bare-metal stent(s)
  CSA = cross-sectional area
  DES = drug-eluting stent(s)
  IVUS = intravascular ultrasound
  MSA = minimal stent area
  MSD = minimal stent diameter
  PCI = percutaneous coronary intervention
  RCT = randomized clinical trial
  SES = sirolimus-eluting stent(s)
  TVR = target vessel revascularization


Optimization of stent deployment during percutaneous coronary intervention (PCI) is a key element to obtain most favorable immediate and long-term results. Since the introduction of balloon-expandable bare-metal stents (BMS) in common practice, the need was recognized for adequate stent expansion to avoid suboptimal stent deployment and reduce the incidence of target vessel revascularization (TVR) (1,2).

Early-generation stents were delivered with a compliant balloon and systematically required further post-dilation with a noncompliant balloon at higher pressures to obtain an adequate lumen dilation and complete apposition of stent struts to the vessel wall (3–6). Notwithstanding the successive introduction of stent delivery systems based on a semi-compliant balloon allowing stent deployment at higher pressures (≥14 atm), post-dilation with a noncompliant balloon or using a larger balloon is still demanded to attain the principle "media to media" stent expansion (7–12).

After the introduction of drug-eluting stents (DES) that dramatically improved restenosis and TVR, the importance of optimal stent deployment was initially underestimated, leading to less use of balloon post-dilation. In the major randomized clinical trials (RCTs) testing DES, post-dilation was not routinely performed (13–15). Nevertheless, despite the lack of evidence from RCTs, observational data continue to support the use of adjunctive balloon post-dilation after deployment of DES in the great majority of patients (16–21). Indeed, the current stent delivery systems of DES are similar to if not the same as in BMS, and the risk of suboptimal stent expansion is still high. Importantly, suboptimal or incomplete stent expansion, especially with DES with polymeric coating, not only might be associated with increased restenosis and TVR, but also might predispose to stent thrombosis (21–25).

The aim of this paper is to review all the possible causes and consequences of a suboptimal stent expansion and assess the possible role of adequate high-pressure balloon post-dilation to prevent it in the DES era.


    Diagnosis and Magnitude of Suboptimal Stent Deployment: Role of Intravascular Ultrasound (IVUS)
 Top
 Abstract
 Diagnosis and Magnitude of...
 Determinants of Suboptimal Stent...
 High-Pressure Stent Deployment...
 Compliant Versus Noncompliant...
 Benefits of a Noncompliant...
 Conclusions
 REFERENCES
 
Because stent underexpansion is poorly recognized by angiography, the real incidence of suboptimal stent deployment is likely to be underestimated. Indeed, it has been observed that discrepancies exist between angiographically-defined and IVUS-defined optimal stent deployment regardless of the stent implanted, with the IVUS success rate raging from 13% to 70% despite successful angiographic results (1,18,26–28). A comparison example between angiographic and IVUS results before and after high-pressure stent post-dilation is shown in Figure 1.


Figure 1
View larger version (178K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 1 Comparison of Angiographic and IVUS Findings Before and After High-Pressure Stent Post-Dilation

Comparison of angiographic and intravascular ultrasound (IVUS) results after high-pressure stent deployment (A) and after post-dilation with a noncompliant balloon of the same size (B). Note that by angiography, no clear difference can be seen before and after the post-dilation. Conversely IVUS demonstrates a significant increase in lumen cross-sectional area (CSA) after a noncompliant balloon post-dilation.

 
An IVUS analysis is more accurate than angiography in determining in-stent dimension and is able to better detect subtle findings such as incomplete apposition and dissection at the stent edges (29–33). Several trials in which stent expansion was assessed by IVUS, such as the STRUT (Stent Treatment Region assessed by Ultrasound Tomography) (34), CRUISE (Can Routine Ultrasound Influence Stent Expansion) (7), and AVID (Angiography Versus Intravascular Ultrasound Directed Coronary Stent Placement) (35) studies, showed an incidence of post-procedural incomplete BMS deployment raging from 4% to 22% (36). Similarly, a high rate of inadequate expansion, ranging between 24% and 28%, has also been observed with current DES (18).

Moreover, IVUS measurements have provided powerful predictive information regarding stent thrombosis, restenosis, and TVR. In particular, minimal stent area (MSA) and minimal stent diameter (MSD) measured by IVUS at the end of the procedure are the strongest predictors of TVR after BMS implantation in numerous studies (37–41). Therefore, IVUS should be considered the gold standard to verify the final result of a PCI procedure and should be recommended in those situations with increased risk of suboptimal stent deployment. Unfortunately this statement, even if it seems reasonable, is so far supported only by "personal opinion" (Level of Evidence C).

One of the main problems in this field is the lack of a uniform and accepted definition of optimal stent expansion by IVUS. Whereas angiographic optimal stent expansion can be qualified as 0% residual stenosis, such a clear definition is not present in the IVUS published data.

The first study that prospectively evaluated the effect of optimal stent expansion according to specific IVUS definitions was the MUSIC (Multicenter Ultrasound Stenting in Coronaries) study (2). Table 1 shows the IVUS criteria introduced in the MUSIC study to define "optimal" stent expansion. Subsequently, other studies assessing the role of IVUS-guided stent delivery adopted slightly different criteria, which are summarized in Table 2.


View this table:
[in this window]
[in a new window]

 
Table 1 Optimal Stent Expansion Criteria Adopted in the MUSIC Study
 

View this table:
[in this window]
[in a new window]

 
Table 2 Optimal Stent Expansion Criteria Adopted in the Studies Assessing the Role of IVUS-Guided Post-Dilation
 
In our view, the main limitation present in all criteria proposed is that none of them takes the advantage given by positive remodeling at the lesion site. The presence of positive remodeling at the lesion site allows the dilation of the stent to a size larger than the distal lumen. This advantage is particularly important when using long stents, where it is useful to size each segment as close as possible to the real dimensions of the vessel and not just to an extrapolated target taken from 2 references (proximal and distal) that are very far from each other. In such circumstances, criteria using an average between the lumen proximal and distal to the stent, such as in the TULIP (Thrombocyte activity evaluation and effects of Ultrasound guidance in Long Intracoronary stent Placement) study (8), becomes of limited value, owing to the discrepancies between proximal and distal lumen sizes. The closer we are to the proximal or distal end of a long stent, the larger will be the inappropriate value of this target criterion.

The recently launched AVIO (Angiography Versus IVUS Optimisation) study, which has been designed to compare the long-term results with DES with an IVUS-guided strategy versus an angiographic strategy, will use a novel criteria (i.e., a minimal luminal area ≥70% the nominal balloon cross-sectional area [CSA] used to dilate a specific segment of the stent) to define optimal stent implantation. In addition, in this study, post-dilation noncompliant balloon size will be selected according to the average (maximum and minimum diameter) of the media to media diameter at the following points: distal in-stent segment, proximal in-stent segment, and point in-stent of maximal narrowing. Alternatively a lumen CSA >9 mm2 will be considered adequate. In our opinion, these new definitions will overcome the limitations used in previous IVUS-guided RCTs.


    Determinants of Suboptimal Stent Deployment
 Top
 Abstract
 Diagnosis and Magnitude of...
 Determinants of Suboptimal Stent...
 High-Pressure Stent Deployment...
 Compliant Versus Noncompliant...
 Benefits of a Noncompliant...
 Conclusions
 REFERENCES
 
Stent undersizing.   Among the possible reasons for suboptimal stent deployment, the first is certainly the undersizing of the stent delivery balloon related to the target vessel. Indeed, in patients with severe and diffuse target vessel disease, the choice of the correct stent size on the basis of only angiographic evaluation is often difficult and leads very often to a balloon to artery ratio <1 (42,43). Our group clearly demonstrated that assessment of the reference vessel diameter in coronary arteries significantly varies according to the method of measurement used; a difference between IVUS and angiography >1.0 mm was found in 71% of cases with vessel size <2.75 mm and in 49% of cases in patients with vessel size >2.75 mm, respectively (42).

Similarly, the recent practice of direct stenting might increase the risk of stent undersizing, especially when handled by nonexperienced operators and when used to treat severe stenotic lesions preventing the correct estimation of the distal vessel size.

It is worth noting that in case of undersizing of the stent delivery balloon, high-pressure stent deployment, especially with the current semi-compliant balloon, can compensate for the balloon undersizing only in part.

Balloon device compliance and pressure deployment.   Another possible cause of stent underexpansion is strictly related to the compliance of balloons commonly used in delivery systems that are often not adequate to guarantee full stent expansion at nominal pressures (26). Stent manufacturers provide a compliance chart relating balloon deployment pressure and the stent diameter. Nevertheless, compliance charts are based on in vitro measurement (in air or in water), exclusively depending on material properties or design features of the stents themselves (44), but in vivo stent expansion is mainly limited by lesion and vessel compliance. Several IVUS studies found that the real MSD after stent deployment was 20% to 26% less than the unconstrained stent size displayed in the compliance chart on the stent box (9,10,18,20,26,45). These differences were independent of stent manufacturer, length, diameter, and deployment pressure and related to the inherent resistance of dilating a stent within an atherosclerotic artery (26,28).

Notwithstanding these discrepancies, the manufacturers’ compliance charts are routinely used by interventional cardiologists to optimize stent diameter according to inflation pressure during PCI. Thus, in clinical practice, the pressure level at which the stent is deployed probably represents the critical determinant of final stent expansion and apposition to the vessel wall. The importance of this issue has not changed with the new stent delivery systems based on semi-compliant balloons, and DES showed the same properties as the corresponding BMS platforms (46). Therefore, high-pressure stent deployment is still strongly recommended to obtain full expansion of both BMS and DES. In this context, IVUS analysis might have a role to check whether the pressures used have really fulfilled the job to optimally deploy the stent.

Plaque and vessel compliance.   In IVUS studies, arterial expansion seemed to be the primary mechanism of lumen enlargement after stenting, accounting for approximately 70% of luminal gain, whereas the relative contribution of plaque reduction ranged between 6% and 34% (47–49). Therefore, the presence of calcium or fibrosis that impairs distensibility of the vessel wall and the presence of high plaque burden behind the stent might represent major limitations to complete stent opening (50).

In this context, the negative influence of plaque burden on adequacy of deployment has been demonstrated by Yoon et al. (28) in a single center study on consecutive patients undergoing stent implantation with IVUS examination. Nevertheless, in this study, the effect of stent post-dilation has not been assessed. Similarly, calcified vessels affect final stent lumen area, preventing complete expansion even when higher pressures or larger balloons are applied (51). In this situation, the use of high-pressure balloon inflations determines vessel overexpansion at noncalcified segments rather than compression of the calcific plaque. The net result is that a significant portion of stent remains underexpanded and asymmetric, which in turn probably explains the higher rate of restenosis found in this type of lesion (39,52).

Therefore, in these situations, when ablative or atherectomy techniques are not feasible, the use of a noncompliant balloon post-dilation represents a good compromise to achieve good stent expansion and symmetry without increasing risk of dissection or rupture of the vessel. In studies in which IVUS analysis after post-dilation was accomplished, including the POSTIT (Postdilatation Clinical Comparative Study) trial (10), none of the baseline clinical or angiographic variables seemed able to predict the final MSA or MSD after stenting. Similarly, neither quantitative IVUS lesion measurements nor qualitative IVUS assessment of plaque morphology could predict stent expansion (18). These observations would suggest that the impact of plaque and vessel compliance on the final stent expansion can be limited by an appropriate use of post-dilation.

Lesion characteristics.   Specific lesion subsets are associated with a lower success rate and required more care and tools to obtain an optimal stent deployment. Lesion subsets in which adequate post-dilation should be carefully considered are summarized in Table 3.


View this table:
[in this window]
[in a new window]

 
Table 3 List of Conditions Associated With an Increased Risk of Suboptimal Stent Deployment
 
Bifurcation lesions
Bifurcation treatment is associated with a high incidence of nonuniform stent expansion in the side branch, resulting in a higher TVR rate (53). Indeed, the lateral opening of the stent in the main branch to gain access to the side branch causes strut deformation and malapposition (54–57). In this context, several studies showed that in bifurcation lesions, especially in the case of both branches stenting, final post-dilation with a kissing balloon is associated with more favorable long-term outcome, reducing the restenosis rate of the side branch and the need for TLR (57–59). Thus, kissing balloon dilation at the end of the procedure is mandatory for bifurcations, but balloon diameters and inflation pressures are yet to be defined to uniformly expand stent struts in both branches. In this context, the use of adequately sized noncompliant balloons at truly high pressures might represent a good compromise between safety and efficacy.

Long lesions
Although lesion length does not influence stent expansion per se (44), the RENEWAL (Randomised Trial of Endoluminal Reconstruction Comparing the NIR Stent and Wallstent in Angioplasty of Long Segment Coronary Disease) and MUSIC trials showed that the high restenosis rates with long stents might be related to a suboptimal stent deployment (2,60). Treatment of long lesions, especially when more than 1 stent is required or when long stents are implanted, increases the risk of size mismatch between the proximal and distal portion of the target vessel. Indeed, in such cases, the stent is usually sized for distal reference diameter and results undersized for the proximal reference diameter. Furthermore, the presence of a double stent struts layer could reduce vessel compliance and can produce an incomplete stent apposition beneath the overlap. For these reasons, a systematic post-dilation of the proximal portion of a long stent and of the overlapping zone with high-pressure inflations or a larger balloon is strongly recommended.

Small vessels
In clinical practice, many interventional cardiologists generally avoid implantation of stents with a diameter size smaller than 2.5 mm. Thus, for small vessels it is common practice to deploy stents that are slightly larger but at lower pressures to avoid excessive vessel overstretch. This habit results in very high incidence of suboptimal stent deployment and an increased incidence of TVR (20). As previously described, in this context it is fundamental to assess whether the vessel is really a small vessel or just appears as such at angiography. Nitroglycerine administration, IVUS evaluation, and bearing in mind that the left anterior descending artery is rarely a small vessel unless in its distal segment are important concepts. If the vessel’s small size were confirmed, post-dilation at higher pressures with a 2.5-mm noncompliant balloon would represent an indispensable solution to achieve the largest final lumen stent to reduce the risk of stent thrombosis and TVR.

In-stent restenosis
Although DES use has also provided encouraging results in this setting, stent underexpansion is the main predictor of in-stent restenosis, ranging between 20% and 40% of the cases (61–66). Interestingly, Blackman et al. (67) showed that a noncompliant balloon post-dilation is necessary during treatment of in-stent restenosis to achieve luminal gain through further expansion of the original stent, which cannot be obtained with a second stent deployment alone. In this study, after post-dilation with a noncompliant balloon, the number of patients with optimal stent expansion according to the IVUS criteria increased from 33% to 67%. Thus, this study also demonstrated that, with post-dilation with noncompliant balloons, significant stent re-expansion can be obtained during repeat angioplasty irrespective of the grade of initial stent underexpansion.


    High-Pressure Stent Deployment Versus Noncompliant Balloon Post-Dilation
 Top
 Abstract
 Diagnosis and Magnitude of...
 Determinants of Suboptimal Stent...
 High-Pressure Stent Deployment...
 Compliant Versus Noncompliant...
 Benefits of a Noncompliant...
 Conclusions
 REFERENCES
 
Several studies (3,6) demonstrated that supranominal pressures are frequently not enough to offset the high impedance of diseased artery. In particular Bermejo et al. (6) showed that, despite high-pressure stent deployment, only 55% of achievable acute lumen gain was effectively obtained, suggesting that plaque characteristics and vessel resistance often cause inadequate expansion of semi-compliant balloons. Unfortunately, there is no guarantee that simple high-pressure post-dilation with a noncompliant balloon will result in final optimal stent deployment.

Consequently, in the subsequent years, the use of IVUS to optimize stent implantation became common practice, and several studies evaluated its role in PCI outcome. The MUSIC study was the first large study that assessed the role of IVUS-guided post-dilation to optimize stent deployment, demonstrating its favorable impact on immediate and 6-month clinical and angiographic results (2).

Similarly, the POSTIT (10), CRUISE (7), and TULIP (8) studies clearly showed that systematic IVUS-guided post-dilation can provide a larger minimal stent area and how this might translate into better long-term outcome. In particular, the multicenter POSTIT trial (10) showed that adjunctive a noncompliant balloon post-dilation can double the frequency of achieving optimal stent deployment when compared with high delivery pressures. Furthermore, the CRUISE study demonstrated the clinical relevance of this approach, showing that an increase of 14% of the final MSA obtained in the IVUS-guided group resulted in a 44% composite relative reduction in TVR (7).

In contrast, other contemporary studies, such as the RESIST (REStenosis after IVUS-guided STenting) (68), OPTICUS (Optimization with ICUS To Reduce Stent Restenosis) (69), SIPS (Strategy for Intracoronary ultrasound-guided PTCA and Stenting) (70), and PRESTO (Prevention of Restenosis with Tranilast and its Outcomes) (71) studies, did not confirm the benefit derived from IVUS guidance for post-dilation on long-term results.

Although there is a trend toward a benefit in TLR favoring IVUS-guided coronary stent implantation, especially in high-risk lesion subtypes (e.g., saphenous vein grafts, long lesions), the effect on long-term death and nonfatal myocardial infarction is neutral (72).

The choice of more conservative IVUS criteria and the consequent less aggressive stent post-dilation (Table 2), the heterogeneity with respect to the rate of patients fulfilling the pre-specified IVUS criteria at the end of procedure, and the underpowered design in most negative studies constitutes possible explanation of these conflicting results. Nevertheless, apart from their specific characteristics and results, all of the studies cited in the preceding text undoubtedly demonstrated that an adequate post-dilation is required to assure an optimal stent deployment in most of the lesions, independently from the original pressure of stent deployment.


    Compliant Versus Noncompliant Balloon Post-Dilation
 Top
 Abstract
 Diagnosis and Magnitude of...
 Determinants of Suboptimal Stent...
 High-Pressure Stent Deployment...
 Compliant Versus Noncompliant...
 Benefits of a Noncompliant...
 Conclusions
 REFERENCES
 
For safety and deliverability reasons, most of the stent delivery systems are currently based on a semi-compliant balloon device. The compliant nature of these balloons causes significant deformation of profile and volume with increases in pressures, resulting in stretching of the balloon itself. Thus, although current semi-compliant balloons are precisely matched with stent length with very short balloon shoulders and assure a uniform diameter expansion along the balloon length, the risk of vessel stretch and edges injury is important at high pressures. Similarly, this technical aspect must be carefully considered during high-pressure stent post-dilation with semi-compliant balloons.

Conversely, noncompliant balloons have little change in volume, even at high pressures concentrating dilating force at the lesion site (73). Indeed, bench tests and clinical studies have shown that noncompliant balloons exert more dilating force against a lesion or a stent than compliant balloons for a given balloon size and inflation pressure. Thus, post-dilation with noncompliant balloons actually results in a significant improvement of MSA compared with the current semi-compliant stent deployment balloons. Especially, the use of a noncompliant balloon post-dilation avoids using a balloon larger than the reference vessel and also allows the application of very high pressures (just below the rated burst pressure) in a safe manner. The main difference between compliant and noncompliant balloons during high-pressure inflation is shown in Figure 2.


Figure 2
View larger version (136K):
[in this window]
[in a new window]
[Download PPT slide]
 
Figure 2 Difference Between Compliant and Noncompliant Balloons During High-Pressure Inflation

Bench test showing the different profile between a noncompliant balloon (A) and a semi-compliant stent delivery balloon (B) during high-pressure (>14 atm) inflation. The semi-compliant balloon demonstrates a "dog bone" effect at the edge of the cylinder that can damage the vessel wall in vivo.

 
These features can be important to prevent unwanted and potentially severe complications caused by over-dilation or uncontrolled fast stretch of the vessel wall. Using a compliant balloon at high pressures, for example, in calcified or very stiff lesions might cause dissection at stent edges and, in small vessels, might facilitate coronary rupture. Several animal and human studies suggested that aggressive stent inflation with high pressures, causing deeper injury of the vessel wall with rupture of the intima or media (74–76), might result in a long-term inflammatory response with a greater neointimal proliferative response and an increased restenosis rate (77,78). Moreover, a SIRIUS (SIRolImUS-Eluting Stent in De Novo Native Coronary Lesions) IVUS substudy (79) suggested that more injury to the contiguous vasoelastic normal wall, coupled with a drug that delays the healing process, could contribute to late stent malapposition owing to focal positive vessel remodeling.

These considerations should lead to more common use of a noncompliant balloon for post-dilation at high pressure.


    Benefits of a Noncompliant Balloon Post-Dilation After DES Implantation
 Top
 Abstract
 Diagnosis and Magnitude of...
 Determinants of Suboptimal Stent...
 High-Pressure Stent Deployment...
 Compliant Versus Noncompliant...
 Benefits of a Noncompliant...
 Conclusions
 REFERENCES
 
The main consequences of stent underexpansion are represented by higher rate of restenosis and related necessity of repeated revascularization at long-term follow-up. Moreover, with DES, potential increase in the risk of stent thrombosis exists (80,81).

In-stent restenosis.   Several studies demonstrated the importance of complete strut apposition, luminal CSA, and concentricity of the stent implanted (29,82). According to these data, IVUS measurement of MSA is the single most powerful predictor of long-term patency and clinical outcome with an inverse relation between post-procedural MSA and angiographic restenosis and between MSA and TVR (4,37,38,40,83). In the TULIP study (8), for example, restenosis rate was 23% in the group guided by IVUS versus 45% in the angiography group, whereas the CRUISE trial (7) demonstrated a relative TVR reduction of 44% when stent underexpansion was corrected by IVUS.

With the advent of DES, determining a lower rate of angiographic restenosis and clinical TVR, the need for adjunctive balloon post-dilation was considered less important. Moreover, the possible role of vessel trauma at the stent margins owing to pre- and post-dilation in determining sirolimus-eluting stent (SES) edge restenosis in the SIRIUS trial (14) led to a more extensive use of direct stenting, whereas post-dilation was used only if strictly required by suboptimal angiographic stent placement (84). In particular, a comparison study (85) between SIRIUS-US (pre- and post-dilation) and E-SIRIUS (direct stenting and nonmandatory post-dilation) IVUS results showed that, although the less aggressive stent implantation in E-SIRIUS resulted in a relatively lower stent expansion, no detrimental effects were observed in major adverse cardiac events.

Subsequently, serial IVUS analyses from the SIRIUS trial showed that, whereas a smaller MSA might be acceptable after DES, underexpansion can result in restenosis (21). This finding was confirmed in other studies in which an incidence of 60% to 70% stent underexpansion determining an MSA <5.0 mm2 was found in most of the cases of SES failures (86–88). Moreover, specific IVUS analysis of DES failure demonstrated a correlation between non-uniform stent strut distribution and maximal neointimal hyperplasia, suggesting local drug underdosing as mainly responsible for this phenomenon (16,89).

In conclusion, observational data suggest that stent underexpansion might be one of the most important causes of DES failure, advocating that, once neointimal hyperplasia is suppressed, the optimum stent deployment is still fundamental. Therefore, post-dilation with noncompliant balloons at high pressures, improving final MSA and MSD, might greatly increase the frequency of optimum DES deployment and actually lead to reduction of restenosis and TVR rates. To specifically address the exact role of high-pressure noncompliant post-dilation strategy, randomized clinical studies are warranted in the DES era.

Stent thrombosis.   Stent thrombosis remains a potential devastating complication in patients undergoing PCI. In the current era of high-pressure stent implantation, the occurrence of stent thrombosis with BMS ranges from 1% to 2% (36,90). Conversely, the rate of DES thrombosis ranges between 0.4% to 0.6% (14,15) of the randomized trials and 1.3% to 4.9% of "real world" registries (24,80).

The major post-procedural predictors of thrombosis with both BMS and DES are MSA and suboptimal stent expansion (23,36,91); in particular, stent underexpansion, resulting in abnormal shear stress, might explain as much as 80% of those events (22,92).

In a large registry of 7,484 consecutive patients treated with BMS, only 22% of patients that experienced subacute stent thrombosis had an optimum PCI result as assessed by IVUS. In this study, the IVUS analyses of thrombosed stents revealed an inadequate lumen dilation (final lumen <80% reference lumen) in 78% of cases, edge dissection in 17%, malapposition in 9%, and plaque prolapse in 4% (92).

The importance of complete stent expansion to prevent stent thrombosis might be more relevant with DES. Indeed, decreased endothelialization associated with drug-related inhibition of neointimal proliferation might increase the risk of stent thrombosis in the case of suboptimal stent expansion. Fujii et al. (23) showed in a retrospective study that lesions leading to stent thrombosis after successful SES implantation more often have stent underexpansion, smaller MSA, and residual edge stenosis. This evidence has also been confirmed for the late-thrombosis cases of DES (>12 months) (91), but in this case, the pathophysiology mechanism is likely more complex (e.g., late acquired incomplete stent apposition) and not completely understood (93).

In this view, the less care used by operators to obtain an optimal stent deployment and a lower use of post-dilation might represent a possible explanation of the higher rates of stent thrombosis observed with DES. Thus, in absence of conclusive data regarding DES long-term safety, it is reasonable to hypothesize that a return to the "antique" practice of a noncompliant balloon post-dilation could also result in an overall reduction of the risk of stent thrombosis with DES.


    Conclusions
 Top
 Abstract
 Diagnosis and Magnitude of...
 Determinants of Suboptimal Stent...
 High-Pressure Stent Deployment...
 Compliant Versus Noncompliant...
 Benefits of a Noncompliant...
 Conclusions
 REFERENCES
 
In clinical practice, according to the rate of suboptimal stent deployment reported in different studies, a considerable number of patients might benefit from repeat inflations with noncompliant balloons at higher pressures and/or with larger diameter size. Indeed, although incomplete stent expansion was attributed as a defect of the first-generation balloon-expandable stent, several studies demonstrated that adjunctive balloon dilation is still necessary to improve the minimum stent area and the volumetric expansion throughout the stented segment.

Data from the literature suggest that achieving adequate stent expansion during PCI is important to reduce restenosis and the need for TVR, but it might also minimize the risk of stent thrombosis in the DES era. Although there are not enough randomized data to support its use, it seems wise to perform post-dilation with noncompliant balloons at high pressures in the majority of patients undergoing both BMS and DES implantation. Particularly in IVUS-guided procedures, the recommended strategy to achieve an optimal stent deployment should be to select a noncompliant balloon whose size matches the media-to-media IVUS measurement.

Unfortunately, it is not practical and cost-effective to perform post-dilation in all patients undergoing stent implantation. A more reasonable approach would then be to select those situations in which the risk of suboptimal stent delivery is higher, especially when DES are employed.

Randomized controlled trials to assess the role of IVUS-guided optimal stent implantation with DES are warranted.

* Reprint requests and correspondence: Dr. Antonio Colombo, San Raffaele Scientific Institute, Cardiac Catheterization Laboratory, Via Buonarroti 48, Milan, Italy 20145. (Email: colombo{at}emocolumbus.it).

Manuscript received July 23, 2007; revised manuscript received October 16, 2007, accepted October 25, 2007.


    REFERENCES
 Top
 Abstract
 Diagnosis and Magnitude of...
 Determinants of Suboptimal Stent...
 High-Pressure Stent Deployment...
 Compliant Versus Noncompliant...
 Benefits of a Noncompliant...
 Conclusions
 REFERENCES
 

  1. Colombo A, Hall P, Nakamura S, et al. Intracoronary stenting without anticoagulation accomplished with intravascular ultrasound guidance Circulation 1995;15:1676-168891.
  2. de Jaegere P, Mudra H, Figulla H, et al. Intravascular ultrasound-guided optimized stent deployment. Immediate and 6 months clinical and angiographic results from the Multicenter Ultrasound Stenting in Coronaries Study (MUSIC Study). Eur Heart J 1998;19:1214-1223.[Abstract/Free Full Text]
  3. Gorge G, Haude M, Ge J, et al. Intravascular ultrasound after low and high inflation pressure coronary artery stent implantation J Am Coll Cardiol 1995;26:725-730.[Abstract]
  4. Albiero R, Rau T, Schluter M, et al. Comparison of immediate and intermediate-term results of intravascular ultrasound versus angiography-guided Palmaz-Schatz stent implantation in matched lesions Circulation 1997;96:2997-3005.[Abstract/Free Full Text]
  5. Werner GS, Diedrich J, Schunemann S, et al. Additional luminal area gain by intravascular ultrasound guidance after coronary stent implantation with high inflation pressure Int J Card Imaging 1997;13:311-321.[CrossRef][Web of Science][Medline]
  6. Bermejo J, Botas J, Garcia E, et al. Mechanisms of residual lumen stenosis after high-pressure stent implantation: a quantitative coronary angiography and intravascular ultrasound study Circulation 1998;98:112-118.[Abstract/Free Full Text]
  7. Fitzgerald PJ, Oshima A, Hayase M, et al. Final results of the Can Routine Ultrasound Influence Stent Expansion (CRUISE) study Circulation 2000;102:523-530.[Abstract/Free Full Text]
  8. Oemrawsingh PV, Mintz GS, Schalij MJ, et al. Intravascular ultrasound guidance improves angiographic and clinical outcome of stent implantation for long coronary artery stenoses: final results of a randomized comparison with angiographic guidance (TULIP Study) Circulation 2003;107:62-67.[Abstract/Free Full Text]
  9. Takano Y, Yeatman LA, Higgins JR, et al. Optimizing stent expansion with new stent delivery systems J Am Coll Cardiol 2001;38:1622-1627.[Abstract/Free Full Text]
  10. Brodie BR, Cooper C, Jones M, et al. Is adjunctive balloon postdilatation necessary after coronary stent deployment?. Final results from the POSTIT trial. Catheter Cardiovasc Interv 2003;59:184-192.[CrossRef][Web of Science][Medline]
  11. Johansson B, Allared M, Borgencrantz B, et al. Standardized angiographically guided over-dilatation of stents using high pressure technique optimize results without increasing risks J Invasive Cardiol 2002;14:221-226.[Medline]
  12. Hur SH, Kitamura K, Morino Y, et al. Efficacy of postdeployment balloon dilatation for current generation stents as assessed by intravascular ultrasound Am J Cardiol 2001;88:1114-1119.[CrossRef][Web of Science][Medline]
  13. Morice MC, Serruys PW, Sousa JE, et al. A randomized comparison of a sirolimus-eluting stent with a standard stent for coronary revascularization N Engl J Med 2002;346:1773-1780.[Abstract/Free Full Text]
  14. Moses JW, Leon MB, Popma JJ, et al. Sirolimus-eluting stents versus standard stents in patients with stenosis in a native coronary artery N Engl J Med 2003;349:1315-1323.[Abstract/Free Full Text]
  15. Stone GW, Ellis SG, Cox DA, et al. Polymer-based, paclitaxel-eluting stent in patients with coronary artery disease N Engl J Med 2004;350:221-231.[Abstract/Free Full Text]
  16. Takebayashi H, Mintz GS, Carlier SG, et al. Non-uniform strut distribution correlates with more neointimal hyperplasia after sirolimus-eluting stent implantation Circulation 2004;110:3430-3434.[Abstract/Free Full Text]
  17. Lemos PA, Hoye A, Goedhart D, et al. Clinical, angiographic, and procedural predictors of angiographic restenosis after sirolimus-eluting stent implantation in complex patients: an evaluation from the Rapamycin-Eluting Stent Evaluated At Rotterdam Cardiology Hospital (RESEARCH) study Circulation 2004;109:1366-1370.[Abstract/Free Full Text]
  18. de Ribamar Costa Jr. J, Mintz GS, Carlier SG, et al. Intravascular ultrasound assessment of drug-eluting stent expansion Am Heart J 2007;153:297-303.[CrossRef][Web of Science][Medline]
  19. Stone GW, Ellis SG, Cox DA, et al. One-year clinical results with the slow-release, polymer-based, paclitaxel-eluting TAXUS stent: the TAXUS-IV trial Circulation 2004;109:1942-1947.[Abstract/Free Full Text]
  20. Cheneau E, Satler LF, Escolar E, et al. Underexpansion of sirolimus-eluting stents: incidence and relationship to delivery pressure Catheter Cardiovasc Interv 2005;65:222-226.[CrossRef][Web of Science][Medline]
  21. Sonoda S, Morino Y, Ako J, et al. Impact of final stent dimensions on long-term results following sirolimus-eluting stent implantation: serial intravascular ultrasound analysis from the SIRIUS trial J Am Coll Cardiol 2004;43:1959-1963.[Abstract/Free Full Text]
  22. Cutlip DE, Baim DS, Ho KK, et al. Stent thrombosis in the modern era: a pooled analysis of multicenter coronary stent clinical trials Circulation 2001;103:1967-1971.[Abstract/Free Full Text]
  23. Fujii K, Carlier SG, Mintz GS, et al. Stent underexpansion and residual reference segment stenosis are related to stent thrombosis after sirolimus-eluting stent implantation: an intravascular ultrasound study J Am Coll Cardiol 2005;45:995-998.[Abstract/Free Full Text]
  24. Iakovou I, Schmidt T, Bonizzoni E, et al. Incidence, predictors, and outcome of thrombosis after successful implantation of drug-eluting stents JAMA 2005;293:2126-2130.[Abstract/Free Full Text]
  25. Fujii K, Mintz GS, Kobayashi Y, et al. Contribution of stent underexpansion to recurrence after sirolimus-eluting stent implantation for in-stent restenosis Circulation 2004;109:1085-1088.[Abstract/Free Full Text]
  26. de Ribamar Costa Jr. J, Mintz GS, Carlier SG, et al. Intravascular ultrasonic assessment of stent diameters derived from manufacturer’s compliance charts Am J Cardiol 2005;96:74-78.[Web of Science][Medline]
  27. Hanekamp CE, Koolen JJ, Pijls NH, Michels HR, Bonnier HJ. Comparison of quantitative coronary angiography, intravascular ultrasound, and coronary pressure measurement to assess optimum stent deployment Circulation 1999;99:1015-1021.[Abstract/Free Full Text]
  28. Yoon SC, Laskey WK, Assadourian A, et al. Assessment of contemporary stent deployment using intravascular ultrasound Catheter Cardiovasc Interv 2002;57:150-154.[CrossRef][Web of Science][Medline]
  29. Nakamura S, Colombo A, Gaglione A, et al. Intracoronary ultrasound observations during stent implantation Circulation 1994;89:2026-2034.[Free Full Text]
  30. Blasini R, Neumann FJ, Schmitt C, Bokenkamp J, Schomig A. Comparison of angiography and intravascular ultrasound for the assessment of lumen size after coronary stent placement: impact of dilation pressures Cathet Cardiovasc Diagn 1997;42:113-119.[CrossRef][Web of Science][Medline]
  31. Mudra H, Klauss V, Blasini R, et al. Ultrasound guidance of Palmaz-Schatz intracoronary stenting with a combined intravascular ultrasound balloon catheter Circulation 1994;90:1252-1261.[Abstract/Free Full Text]
  32. Laskey WK, Brady ST, Kussmaul WG, et al. Intravascular ultrasonographic assessment of the results of coronary artery stenting Am Heart J 1993;125:1576-1583.[CrossRef][Web of Science][Medline]
  33. Keren G, Pichard AD, Kent KM, Satler LF, Leon MB. Failure or success of complex catheter-based interventional procedures assessed by intravascular ultrasound Am Heart J 1992;123:200-208.[CrossRef][Web of Science][Medline]
  34. Metz JA, Mooney MR, Walter PD, et al. Significance of edge tears in coronary stenting: initial observations from the STRUT registry (abstr) Circulation 1995;92(Suppl):I546.
  35. Russo RJ, Attubato MJ, Davidson CJ, et al. Angiography versus intravascular ultrasound-directed stent placement: final results from AVID (abstr) Circulation 1999;100(Suppl I):I234.
  36. Uren NG, Schwarzacher SP, Metz JA, et al. Predictors and outcomes of stent thrombosis: an intravascular ultrasound registry Eur Heart J 2002;23:124-132.[Abstract/Free Full Text]
  37. Hoffmann R, Mintz GS, Mehran R, et al. Intravascular ultrasound predictors of angiographic restenosis in lesions treated with Palmaz-Schatz stents J Am Coll Cardiol 1998;31:43-49.[Abstract/Free Full Text]
  38. de Feyter PJ, Kay P, Disco C, Serruys PW. Reference chart derived from post-stent-implantation intravascular ultrasound predictors of 6-month expected restenosis on quantitative coronary angiography Circulation 1999;100:1777-1783.[Abstract/Free Full Text]
  39. Kuntz RE, Safian RD, Carrozza JP, Fishman RF, Mansour M, Baim DS. The importance of acute luminal diameter in determining restenosis after coronary atherectomy or stenting Circulation 1992;86:1827-1835.[Abstract/Free Full Text]
  40. Kasaoka S, Tobis JM, Akiyama T, et al. Angiographic and intravascular ultrasound predictors of in-stent restenosis J Am Coll Cardiol 1998;32:1630-1635.[Abstract/Free Full Text]
  41. Moussa I, Moses J, Di Mario C, et al. Does the specific intravascular ultrasound criterion used to optimize stent expansion have an impact on the probability of stent restenosis? Am J Cardiol 1999;83:1012-1017.[CrossRef][Web of Science][Medline]
  42. Briguori C, Tobis J, Nishida T, et al. Discrepancy between angiography and intravascular ultrasound when analysing small coronary arteries Eur Heart J 2002;23:247-254.[Abstract/Free Full Text]
  43. Choi JW, Vardi GM, Meyers SN, et al. Role of intracoronary ultrasound after high-pressure stent implantation Am Heart J 2000;139:643-648.[Web of Science][Medline]
  44. Olbrich T, Williams DO, Doig JC, Murray A. In vivo assessment of coronary artery angioplasty and stent deployment from balloon pressure-volume data Physiol Meas 2006;27:213-223.[CrossRef][Web of Science][Medline]
  45. Hehrlein C, DeVries JJ, Wood TA, et al. Overestimation of stent delivery balloon diameters by manufacturers’ compliance tables: a quantitative coronary analysis of Duet and NIR stent implantation Catheter Cardiovasc Interv 2001;53:474-478.[CrossRef][Web of Science][Medline]
  46. Javaid A, Chu WW, Cheneau E, et al. Comparison of paclitaxel-eluting stent and sirolimus-eluting stent expansion at incremental delivery pressures Cardiovasc Revasc Med 2006;7:208-211.[CrossRef][Medline]
  47. Boschat J, Le Breton H, Commeau P, et al. Is coronary stent deployment and remodeling affected by predilatation?. An intravascular ultrasound randomized study. Int J Cardiovasc Imaging 2002;18:399-404.[CrossRef][Web of Science][Medline]
  48. Ahmed JM, Mintz GS, Weissman NJ, et al. Mechanism of lumen enlargement during intracoronary stent implantation: an intravascular ultrasound study Circulation 2000;102:7-10.[Abstract/Free Full Text]
  49. Suneja R, Nair RN, Reddy KG, Rasheed Q, Sheehan HM, Hodgson JM. Mechanisms of angiographically successful directional coronary atherectomy: evaluation by intracoronary ultrasound and comparison with transluminal coronary angioplasty Am Heart J 1993;126:507-514.[CrossRef][Web of Science][Medline]
  50. Prati F, Di Mario C, Moussa I, et al. In-stent neointimal proliferation correlates with the amount of residual plaque burden outside the stent: an intravascular ultrasound study Circulation 1999;99:1011-1014.[Abstract/Free Full Text]
  51. Albrecht D, Kapers S, Fussl R. Coronary plaque morphology affects stent deployment: assessment by intracoronary ultrasound Cathet Cardiovasc Diagn 1996;38:229-235.[CrossRef][Web of Science][Medline]
  52. Vavuranakis M, Toutouzas K, Stefanadis C, Chrisohou C, Markou D, Toutouzas P. Stent deployment in calcified lesions: can we overcome calcific restraint with high-pressure balloon inflations? Catheter Cardiovasc Interv 2001;52:164-172.[CrossRef][Web of Science][Medline]
  53. Colombo A, Moses JW, Morice MC, et al. Randomized study to evaluate sirolimus-eluting stents implanted at coronary bifurcation lesions Circulation 2004;109:1244-1249.[Abstract/Free Full Text]
  54. Ormiston JA, Webster MW, Ruygrok PN, Stewart JT, White HD, Scott DS. Stent deformation following simulated side-branch dilatation: a comparison of five stent designs Catheter Cardiovasc Interv 1999;47:258-264.[CrossRef][Web of Science][Medline]
  55. Lefevre T, Louvard Y, Morice MC, et al. Stenting of bifurcation lesions: classification, treatments, and results Catheter Cardiovasc Interv 2000;49:274-283.[CrossRef][Web of Science][Medline]
  56. Ormiston JA, Currie E, Webster MW, et al. Drug-eluting stents for coronary bifurcations: insights into the crush technique Catheter Cardiovasc Interv 2004;63:332-336.[CrossRef][Web of Science][Medline]
  57. Costa RA, Mintz GS, Carlier SG, et al. Bifurcation coronary lesions treated with the "crush" technique: an intravascular ultrasound analysis J Am Coll Cardiol 2005;46:599-605.[Abstract/Free Full Text]
  58. Ge L, Airoldi F, Iakovou I, et al. Clinical and angiographic outcome after implantation of drug-eluting stents in bifurcation lesions with the crush stent technique: importance of final kissing balloon post-dilation J Am Coll Cardiol 2005;46:613-620.[Abstract/Free Full Text]
  59. Di Mario C, Morici N, Godino C, et al. Predictors of restenosis after treatment of bifurcational lesions with paclitaxel eluting stents: a multicenter prospective registry of 150 consecutive patients Catheter Cardiovasc Interv 2007;69:416-424.[CrossRef][Web of Science][Medline]
  60. Nageh T, de Belder AJ, Thomas MR, Williams IL, Wainwright RJ. A randomised trial of endoluminal reconstruction comparing the NIR stent and the Wallstent in angioplasty of long segment coronary disease: results of the RENEWAL Study Am Heart J 2001;141:971-976.[CrossRef][Web of Science][Medline]
  61. Sousa JE, Costa MA, Abizaid A, et al. Sirolimus-eluting stent for the treatment of in-stent restenosis: a quantitative coronary angiography and three-dimensional intravascular ultrasound study Circulation 2003;107:24-27.[Abstract/Free Full Text]
  62. Tanabe K, Serruys PW, Grube E, et al. TAXUS III Trial: in-stent restenosis treated with stent-based delivery of paclitaxel incorporated in a slow-release polymer formulation Circulation 2003;107:559-564.[Abstract/Free Full Text]
  63. Degertekin M, Regar E, Tanabe K, et al. Sirolimus-eluting stent for treatment of complex in-stent restenosis: the first clinical experience J Am Coll Cardiol 2003;41:184-189.[Abstract/Free Full Text]
  64. Fujii K, Mintz GS, Kobayashi Y, et al. Contribution of stent underexpansion to recurrence after sirolimus-eluting stent implantation for in-stent restenosis Circulation 2004;109:1085-1088.[Abstract/Free Full Text]
  65. Escolar E, Mintz GS, Canos D, et al. Serial intravascular ultrasound comparison of the extent and distribution of intimal hyperplasia six months after stent implantation for de novo versus in-stent restenosis lesions Am J Cardiol 2005;96:897-900.[CrossRef][Web of Science][Medline]
  66. Bertrand OF, De Larochelliere R, Joyal M, Bonan R, Mongrain R, Tardif JC. Incidence of stent under-deployment as a cause of in-stent restenosis in long stents Int J Cardiovasc Imaging 2004;20:279-284.[CrossRef][Web of Science][Medline]
  67. Blackman DJ, Porto I, Shirodaria C, Channon KM, Banning AP. Usefulness of high-pressure post-dilatation to optimize deployment of drug-eluting stents for the treatment of diffuse in-stent coronary restenosis Am J Cardiol 2004;94:922-925.[CrossRef][Web of Science][Medline]
  68. Schiele F, Meneveau N, Vuillemenot A, et al. Impact of intravascular ultrasound guidance in stent deployment on 6-month restenosis rate: a multicenter, randomized study comparing two strategies—with and without intravascular ultrasound guidance J Am Coll Cardiol 1998;32:320-328.[Abstract/Free Full Text]
  69. Mudra H, di Mario C, de Jaegere P, et al. Randomized comparison of coronary stent implantation under ultrasound or angiographic guidance to reduce stent restenosis (OPTICUS Study) Circulation 2001;104:1343-1349.[Abstract/Free Full Text]
  70. Frey AW, Hodgson JM, Muller C, Bestehorn HP, Roskamm H. Ultrasound-guided strategy for provisional stenting with focal balloon combination catheter: results from the randomized Strategy for Intracoronary Ultrasound-guided PTCA and Stenting (SIPS) trial Circulation 2000;102:2497-2502.[Abstract/Free Full Text]
  71. Orford JL, Denktas AE, Williams BA, et al. Routine intravascular ultrasound scanning guidance of coronary stenting is not associated with improved clinical outcomes Am Heart J 2004;148:501-506.[CrossRef][Web of Science][Medline]
  72. Casella G, Klauss V, Ottani F, Siebert U, Sangiorgio P, Bracchetti D. Impact of intravascular ultrasound-guided stenting on long-term clinical outcome: a meta-analysis of available studies comparing intravascular ultrasound-guided and angiographically guided stenting Catheter Cardiovasc Interv 2003;59:314-321.[CrossRef][Web of Science][Medline]
  73. Abele JE. Balloon catheters and transluminal dilatation: technical considerations Am J Roentgenol 1980;135:901-906.[Abstract]
  74. Schwartz RS, Huber KC, Murphy JG, et al. Restenosis and the proportional neointimal response to coronary artery injury: results in a porcine model J Am Coll Cardiol 1992;19:267-274.[Abstract]
  75. Leung DY, Glagov S, Mathews MB. Cyclic stretching stimulates synthesis of matrix components by arterial smooth muscle cells in vitro Science 1976;191:475-477.[Abstract/Free Full Text]
  76. Shi Y, O’Brien JE, Fard A, Mannion JD, Wang D, Zalewski A. Adventitial myofibroblasts contribute to neointimal formation in injured porcine coronary arteries Circulation 1996;94:1655-1664.[Abstract/Free Full Text]
  77. Uretsky BF, Rosanio S, Lerakis S, et al. A prospective evaluation of angiography-guided coronary stent implantation with high versus very high balloon inflation pressure Am Heart J 2000;140:804-812.[CrossRef][Web of Science][Medline]
  78. Farb A, Sangiorgi G, Carter AJ, et al. Pathology of acute and chronic coronary stenting in humans Circulation 1999;99:44-52.[Abstract/Free Full Text]
  79. Ako J, Morino Y, Honda Y, et al. Late incomplete stent apposition after sirolimus-eluting stent implantation: a serial intravascular ultrasound analysis J Am Coll Cardiol 2005;46:1002-1005.[Abstract/Free Full Text]
  80. Pfisterer M, Brunner-La Rocca HP, Buser PT, et al. Late clinical events after clopidogrel discontinuation may limit the benefit of drug-eluting stents: an observational study of drug-eluting versus bare-metal stents J Am Coll Cardiol 2006;48:2584-2591.[Abstract/Free Full Text]
  81. Kastrati A, Mehilli J, Pache J, et al. Analysis of 14 trials comparing sirolimus-eluting stents with bare-metal stents N Engl J Med 2007;356:1030-1039.[Abstract/Free Full Text]
  82. Castagna MT, Mintz GS, Leiboff BO, et al. The contribution of "mechanical" problems to in-stent restenosis: an intravascular ultrasonographic analysis of 1090 consecutive in-stent restenosis lesions Am Heart J 2001;142:970-974.[CrossRef][Web of Science][Medline]
  83. Mintz GS. Mechanisms, prevention, and treatment of restenosisIn: Mintz GS, editor. Intracoronary Ultrasound. London: Taylor and Francis; 2005. pp. 299-400.
  84. Schofer J, Schluter M, Gershlick AH, et al. Sirolimus-eluting stents for treatment of patients with long atherosclerotic lesions in small coronary arteries: double-blind, randomised controlled trial (E-SIRIUS) Lancet 2003;362:1093-1099.[CrossRef][Web of Science][Medline]
  85. Hoffmann R, Guagliumi G, Musumeci G, et al. Vascular response to sirolimus-eluting stents delivered with a nonaggressive implantation technique: comparison of intravascular ultrasound results from the multicenter, randomized E-SIRIUS, and SIRIUS trials Catheter Cardiovasc Interv 2005;66:499-506.[CrossRef][Web of Science][Medline]
  86. Hong MK, Mintz GS, Lee CW, et al. Intravascular ultrasound predictors of angiographic restenosis after sirolimus-eluting stent implantation Eur Heart J 2006;27:1305-1310.[Abstract/Free Full Text]
  87. Takebayashi H, Kobayashi Y, Mintz GS, et al. Intravascular ultrasound assessment of lesions with target vessel failure after sirolimus-eluting stent implantation Am J Cardiol 2005;95:498-502.[CrossRef][Web of Science][Medline]
  88. Kim SW, Mintz GS, Escolar E, et al. An intravascular ultrasound analysis of the mechanisms of restenosis comparing drug-eluting stents with brachytherapy Am J Cardiol 2006;97:1292-1298.[CrossRef][Web of Science][Medline]
  89. Hwang CW, Wu D, Edelman ER. Physiological transport forces govern drug distribution for stent-based delivery Circulation 2001;104:600-605.[Abstract/Free Full Text]
  90. Moussa I, Di Mario C, Reimers B, Akiyama T, Tobis J, Colombo A. Subacute stent thrombosis in the era of intravascular ultrasound-guided coronary stenting without anticoagulation: frequency, predictors and clinical outcome J Am Coll Cardiol 1997;29:6-12.[Abstract]
  91. Cook S, Wenaweser P, Togni M, et al. Incomplete stent apposition and very late stent thrombosis after drug-eluting stent implantation Circulation 2007;115:2426-2434.[Abstract/Free Full Text]
  92. Cheneau E, Leborgne L, Mintz GS, et al. Predictors of subacute stent thrombosis: results of a systematic intravascular ultrasound study Circulation 2003;108:43-47.[Abstract/Free Full Text]
  93. Siqueira DA, Abizaid AA, Costa Jde R, et al. Late incomplete apposition after drug-eluting stent implantation: incidence and potential for adverse clinical outcomes Eur Heart J 2007;28:1304-1309.[Abstract/Free Full Text]

Related articles in JACC Interventions:

Drug-Eluting Stent Implantation: Technique Matters, More Than Ever
Bernard De Bruyne and Emanuele Barbato
JACC Interventions 2008 1: 32-33. [Full Text]  



This article has been cited by other articles:


Home page
J Am Coll Cardiol IntvHome page
B. De Bruyne and E. Barbato
Drug-eluting stent implantation technique matters, more than ever.
J. Am. Coll. Cardiol. Intv., February 1, 2008; 1(1): 32 - 33.
[Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Figures Only
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Related articles in JACC Interventions
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Romagnoli, E.
Right arrow Articles by Colombo, A.
Right arrow Search for Related Content
PubMed
Right arrow Articles by Romagnoli, E.
Right arrow Articles by Colombo, A.

 
   
 
home link current link search link archive link topics link cardiology careers link