top banner image  

topleft corner image     top right corner image
 


bullet

JACC Homepage JACC Imaging Homepage
Still not a subscriber to JACC Imaging or JACC Interventions?

     top nav image

     

J Am Coll Cardiol Intv, 2009; 2:284-290, doi:10.1016/j.jcin.2008.12.012
© 2009 by the American College of Cardiology Foundation
This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by van der Giessen, W. J.
Right arrow Articles by van Beusekom, H. M.M.
Right arrow Search for Related Content
PubMed
Right arrow Articles by van der Giessen, W. J.
Right arrow Articles by van Beusekom, H. M.M.

Mini-Focus: Stent Technology

Lowering the Dose of Sirolimus, Released From a Nonpolymeric Hydroxyapatite Coated Coronary Stent, Reduces Signs of Delayed Healing

Wim J. van der Giessen, MD, PhD*,{dagger}, Oana Sorop, PhD*, Patrick W. Serruys, MD, PhD*, Ilona Peters-Krabbendam, BSc*, Heleen M.M. van Beusekom, PhD*,*

* Department of Cardiology, Erasmus Medical Center, Rotterdam, the Netherlands
{dagger} Interuniversity Cardiology Institute-KNAW, Utrecht, the Netherlands

* Reprint requests and correspondence: Dr. Heleen van Beusekom, Thoraxcenter, Room Ee2393a or 2355a, Erasmus Medical Center, P.O. Box 2040, 3000 CA Rotterdam, the Netherlands (Email: h.vanbeusekom{at}erasmusmc.nl).

Objectives: The aim of this study was to compare efficacy of low- and high-dose sirolimus release (25, 40, or 100 µg) from hydroxyapatite (HAp) with Cypher (Cordis, Johnson & Johnson, Warren, New Jersey) (111 µg sirolimus) in porcine coronary arteries.

Background: Polymer-based sirolimus-eluting stents such as Cypher interfere with vascular healing, probably due to the permanent presence of the polymer coating and the high sirolimus dose. The use of low-dose sirolimus and inert nonpolymeric but biodegradable coatings such as HAp might be more appropriate.

Methods: Stents (n = 68) were implanted, guided by quantitative coronary angiography. All swine received clopidogrel and acetylsalicylic acid during 28 days follow-up. Safety of the coating in absence of drugs was studied by comparing HAp with and without a lipid-based release regulating layer (HApR) with bare-metal stents. Efficacy was studied by comparing the release of 25, 40, and 100 µg sirolimus with Cypher.

Results: The safety study (without drug) revealed no differences in neointimal thickening in response to HAp and HApR with complete healing in all groups. Dose response analysis showed that neointimal thickening was similar in all groups regardless of sirolimus dose, with a normal appearance of the endothelium. There was, however, a dose-dependent increase in fibrinoid (p = 0.028), considered to be a marker of delayed healing. The Cypher stent induced the highest amount of fibrinoid.

Conclusions: Reducing the dose of sirolimus eluting from a biocompatible HAp coated stent reduces signs of delayed vascular healing, without affecting neointimal hyperplasia.

Key Words: animal model • coronary • drug eluting stent • nonpolymeric coatings • PCI

Abbreviations and Acronyms
  ASA = acetylsalicylic acid
  DES = drug-eluting stent(s)
  HAp = hydroxyapatite
  HApR = hydroxyapatite including drug release regulating lipid layer
  QCA = quantitative coronary angiography






 
   
 